Buckleys 6004-0079 Operating Instructions Manual Download Page 6

Instruction Manual

 

6

 

 

 

 

Unpacking 

 

Check that the following items are in the received package: 

 

Quantity  Item

 

Surface Display Pro’ 

Calibration lead (BCM to SDU) 

Mains power lead 

Probe lead (4mm plug to clip) 

Manual 

 

If any of these items are missing please contact your supplier immediately. The 
packaging should be retained for the future to be used if the unit needs to be 
sent back for repair, or needs to be stored. 

 

Operation 

 

Buckleys’  Surface Display Pro’ (SDU)  provides  a  surface  readout  of  the 
corrosion potential voltage  reading  being  taken  by  a  diver  operated 
BathyCorrometer

®  

Pro’. To use the Surface Display Unit, follow the steps below. 

• Ensure that the SDU has been calibrated before use. In general, it is best to 

calibrate the SDU and BathyCorrometer

®  

Pro’ before the commencement of 

each corrosion survey. 

• Connect the SDU to the BathyCorrometer

®  

Pro’ using  the two core, water 

resistant cable  that  is  to  be  used  during the  survey  (supplied  separately). 
Connect the  SDU  to  the  mains  power  lead  ensuring  that  the mains  input 
voltage selector on the rear of the SDU is set correctly. Switch on the mains 
power supply. 

•  Before beginning a  corrosion survey  it  is  a  good idea  to  carry out  a final 

check of the SDU and BathyCorrometer

®  

Pro’. This can be done using a zinc 

test block immersed in a standard salinity solution (3% at 25°C). When the 
BathyCorrometer

®

’s  steel  tip  is  held  against  the  test  block  in  the  saline 

solution the meter reading (approx. 1.00 - 1.08V) should be repeated on the 
SDU. If this is not the case the calibration procedure should be repeated. 

 

When you are satisfied that all the checks have been made and passed, the 
BCM and SDU are ready for use. 

 

Summary of Contents for 6004-0079

Page 1: ...ss MANUFACTURERS OF SPECIALIST TEST EQUIPMENT Surface Display Pro Operating Instructions English Version Please read carefully before use...

Page 2: ......

Page 3: ...osal Information 8 EC Declaration of Conformity 9 UKCA Declaration of Conformity 10 Contact Details 11 Attention Misuse or failure to comply with the guidelines outlined in this manual may impair the...

Page 4: ...urface Display Pro allowing quick and easy verification Note The voltage is now displayed as a negative voltage rather than positive as with early SDUs and Roxby Buckleys BathyCorrometer s This has be...

Page 5: ...r However as with any mains powered device it is advisable to take a few basic precautions to minimize any risk still further Under no circumstances should the case of the SDU be opened If the unit sh...

Page 6: ...the SDU and BathyCorrometer Pro before the commencement of each corrosion survey Connect the SDU to the BathyCorrometer Pro using the two core water resistant cable that is to be used during the surv...

Page 7: ...upplied Plug the supplied mains power lead into the SDU checking that the mains input voltage selector is in the correct position do not connect the mains supply at this stage Use a small flat bladed...

Page 8: ...Buckleys Surface Display Pro has been designed so that no maintenance is required by the user The Surface Display Pro should be inspected from time to time to check that the case calibration leads an...

Page 9: ...ormity We Buckleys UVRAL Ltd As manufacturer of the apparatus listed declare that the product Surface Display Pro Unit Is manufactured in conformity with European directives 2014 30 EU 2014 35 EU Auth...

Page 10: ...cturer of the apparatus listed declare that the product Surface Display Pro Unit Is manufactured in conformity with the following UK legislation Electronic Compatibility Regulations 2016 The Electrica...

Page 11: ...t details Manufacturer Buckleys UVRAL Ltd Address Buckleys House Unit G Concept Court Shearway Business Park Shearway Road Folkestone Kent CT19 4RG UK Tel 44 0 1303 278888 Website www buckleysinternat...

Page 12: ...ating to your product Be the first to hear about our new products We strive to improve the quality of our products and service Registering your product helps us monitor overall quality of our products...

Reviews: