Syncore
Platform
21
If the UP-key of the time controller is hold until“to 1” appears,
the equipment starts the program again using the 1st pro-
gram step and the program will repeat indefinitely in a cyclic
manner. This step can be interrupted by the user by pressing
the „STOP“-Key.
7. Checking the program:
Press the „PROGRAM“ button repeatedly and check
the para meters in all steps of the program. The display
jumps back to step 1 at the end of the program.
8. Starting the program:
Pressing once on the „START“ button
starts the pro-
gram. The program always starts with the first step, no
matter what step is shown on the display.
Comments:
A program cannot be started unless the programming mode
has been activated (which is shown by the indication of a
step in the display
), If only a dot, and no step, appears in
display field
, press once on the „PROGRAM“ button to ac-
tivate the program mode. The program can then be started.
The program mode may be aborted any time by pressing
twice on the „STOP“ button. The temperature and time para-
meters entered previously are stored. A program remains
stored even after the unit is switched off.
Once the speed for the vortex movement is set, it remains
constant for all program steps. It can, however, be changed
manually at any time.
9. Changing parameters while a program or a step in it is in
progress: Activate the set-point temperature and/or the
set-point time as described in steps 2 and 3 and set the
new set-point temperature and/or set-point time. After
a delay of 3 seconds, the unit takes over the new set-
point parameters. These new set-point parameters are,
however, not stored after completion of the program.
10. Ending the program:
A program can be aborted any time by pressing once on
the „STOP“ button. This stops the temperature control
for the base plate and/or the heater.
Quit the programming mode by pressing the „STOP“
twice.