BÜCHI Labortechnik AG
Operation | 6
Operation Manual ProxiMate™
69/94
6.15
Carrying out a measurement
NOTE
Barcode / Sample ID / Order / Note
Barcode, Sample ID, Order and Note for a sample can be edited during the measure-
ment process at any time.
4
See Chapter "Insert sample ID", page 31
4
See Chapter "Insert barcode", page 31
4
See Chapter "Insert note", page 32
4
See Chapter "Insert order", page 32
NOTE
Returning the test sample in the production can cause contamination.
4
Do not return test sample in the production.
NOTE
Contaminated or defective sample carrier
A contaminated or defective sample carrier causes measuring errors.
4
Do not use defective sample carriers.
4
Make sure that the sample carrier is clean.
NOTE
Wrong measurement results due to overfilled petri dishes.
The sample material can fall out from overfilled Petri dishes. This sample material
can lead to an accumulation of material on the measurement window leading to in-
correct measurement results.
4
Do not overfill petri dishes.
6.15.1
Preparing the instrument
Time required:
up to 30 min
Precondition:
R
All commissioning operations have been completed. See Chapter 5 "Installation",
page 21
u
Tap the
On/Off
master switch.
ð
The system starts up.
ð
The display highlights the status bar yellow.
ð
Once the startup phase is complete the status bar changes from yellow to
black.
6.15.2
Starting measurement
Navigation path
➔
Precondition:
R
The instrument is prepared. See Chapter 6.15.1 "Preparing the instrument",
page 69
u
Navigate to the
[Application]
menu via the navigation path.