6 Operation
41
Melting Point M-565 Operation Manual, Version F
When a measurement is done, the last method parameters are always retained in the instrument. To
start the same method, just press
START
.
NOTE
Results are stored in memory until a new measurement is started or the instrument is switched off.
6 .3 .5
Printout
a
b
c
d
e
f
Printout
a
Instrument type
b
Determination parameters
c
Automatically detected results
d
Manual results
e
General information
f
Signature of person making the de-
termination
Boiling temp. = Temperature sample boiled.
Boiling point = Boiling temp. corrected according entered baro-
metric pressure.
If a barometric pressure of 1013 mbar is entered
boiling temperature and boiling point results are
equal.