Err7
The CUFF Pressure Range was
inflated above 295mmHg
when inflating the CUFF manually.
Do not inflate the CUFF Pressure Range above
295mmHg (39.3kPa), Refer to the clause
“BLOOD
PRESSURE MEASUREMENT METHOD”
.
Err8
The Systolic was measured out of the
range
1. Check whether the CUFF joint is loose.
2. The SYSTOLIC Blood Pressure is over
measurement range, please contact your
doctor and follow his or her instructions.
Err9
The Diastolic was measured out of the
range.
1. Check whether the CUFF joint is loose.
2. The DIASTOLIC Blood Pressure is over
measurement range, please contact your doctor
and follow his or her instructions.
Battery is worn.
Please recharge in time.
If the problem is still not resolved, please contact with the dealer or our customer service
department, so as to obtain further assistance.
12 NOTICE
1) The interval between two measurements shall be at least 2 to 3 minutes. The waiting between two
measurements is mainly to get the artery back to the state before the blood pressure is measured.
2) When in repeated measurement, because the Arm blood flow may not be smooth, you may not
measure the correct blood pressure. Please conduct measurement after the blood flow is smooth.
3)
When the LCD display an Error code “Err”, the measurement cannot be conducted correctly, please
refer to the error and fault instructions in the next page.
4) When a measured person has an arrhythmia, the measurement value of blood pressure and pulse rate
may not be accurate enough, so for patients with arrhythmia, it is advised to consult a doctor to make
a professional diagnosis.
5) When the CUFF is too loose or too tight, it will affect the accuracy of measurement.
6) Measurement posture and physical condition, etc. will affect the measurement results. Please
understand the knowledge of blood pressure, and eliminate all kinds of interference factors.
7) If the CUFF not provided by the manufacturer to replace the original one, it may lead to measurement
error. Please do not use the CUFF other than supplied by the manufacturer.
8) Please do not use the blood pressure monitor with electrical appliances that emit electromagnetic
interference, it may affect the accuracy.
9) The key component list of the machine includes main control chip, sensor, pump, electromagnetic valve,
uniform speed air release valve and CUFF. Parts of the manufacturer’s models and specifications
specified in the list must be used; otherwise it may lead to problems of safety or accuracy.