10
3) BatteryCharge Indicator
The power insufficiency battery icon (
or
)will flicker to reminder user the monitor battery under
charging.
The Power fully battery icon (
) is fixed to display after the monitor battery is charged.
WARNING
Do not plug or unplug the power cord into the electrical outlet with wet hands.
The Blood Pressure Monitor cannot measure when charging.
CAUTION
Do not overload power outlets. Plug the device into appropriate voltage outlet.
The device used to provide power must meet the requirements of IEC 60601-1.
The output of the device used to provide power should be 5V d.c.1A, connect port: USB power port
(for power supply only, no data communication function).
9.3 Blood Pressure Measurement Method
9.3.1 Applying the CUFF
1) Make sure the air plug is securely inserted in the main unit.
2) Remove tight-fitting clothing from your upper arm.
3) Push open the CUFF into a tubular shape, until it is suitable for the arm to get in.
4) Place the CUFF around a bare Arm 2-3cm above the Arm joint on the palm side of the Arm.
5) Sit in a chair with your arms on the table in front of you, palms facing up, the display faces yours eyes.