20
high impedance) like fuses to protect the user and the instrument. Most measuring
functions through this terminal will then be open circuit. The series input protection
resistors and the spark-gaps (or varistors) should then be replaced by qualified
technician. Refer to the LIMITED WARRANTY section for obtaining warranty or
repairing service.
Battery and Fuse replacement
Battery use: Four 1.5V AA battery (IEC LR6)
Fuses use:
Fuse (F2) for
/
mA input:
0.4A/1000Vac & Vdc, IR 30kA or better, FF fuse; Dimension: 6 x 32 mm
Fuse (F3) for Earth Continuity Test input:
0.25A/1000Vac & Vdc, IR 30kA or better, FF fuse; Dimension: 6 x 32 mm
Battery and Fuse replacement:
Loosen the screws from the access cover of the case bottom. Lift the access cover.
Replace the batteries or fuse. Re-fasten the screws.