27
55
C)
DC Current
RANGE
Accuracy
Burden Voltage
600.00
A
1) 2)
0.075%+20d
0.2mV/
A
6000.0
A
0.075%+20d
0.2mV/
A
60.000mA
2)
0.075%+20d
2.0mV/mA
600.00mA
0.15%+20d
2.0mV/mA
6.0000A
0.3%+20d
30mV/A
10.000A
3)
0.3%+30d
30mV/A
1)
Specified with Open-circuit-voltage (OCV) of Current-loop-under-test at >100
V.
2)
The meter shows a few negative residue counts when the input is short-circuited, with OCV
at zero volt. It is the nature of the internal protection circuity design, and will not affect
measurement readings at nominal OCVs greater than 100
V in significant measurements.
3)
10A continuous up to ambient 40
o
C only, and is <3 mins on per >15 mins off @ 40
o
C ~
55
o
C; >10A to 20A for <30 seconds on per >15 mins off