Temperature over 165
o
F (74
o
C) at photocell
For gas flames (Flame Rod):
Ignition interference (A flame signal current difference with th e ignition both on and off
greater than 0.5 A indicates the presence of ignition interference)
Insufficient ground (must be at least 4 times the detector area)
Flame lifting off burner head (ground), or not continuously in contact with the flame rod
Temperature in excess of 600
o
F (316
o
C) at the flame electrode insulator causing short
to ground.
Press the
button momentarily to toggle EF Detection feature. T
when it is read y. If it is too sensitive for your applications, press
button momentarily
series of bargraph segments on the display plus variable beep tones.
: An antenna is located along the top
left end of the meter,
which detects electric field surrounding energized live conductors. It is ideal for tracing live
wiring connections, locating wiring breakages and to distinguish between
live and earth
connections.
: For more precise indication of live wires, such as
distinguishing between Live and Ground connections,
use direct contact testing with one
single testprobe via the input terminal COM or V. The COM t erminal (Black) has the best
sensitivity.