Error Descriptions
Daedalus Lock Manual
Spectrospin AG
95
LOCK ErrorOnLCB
Error 4
This is displayed if the CPU, Keyboard or X32 ignore an
error that is detected by the Lock and sent to the CPU.
LOCK Error_Startchar
Error 30
The lock application software is a INTEL Hex file. If the
starting character of any line in this file violates the format
specifications downloading will stop immediately.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
LOCK FFAError
Error 143
Appears if the DSP FFA is not correctly completed within a
certain time period.
Possible Causes:
- Signal ADC_CONV (L-RX) is not available
- ADC (IC9) on the lock receiver is defect.
LOCK GainBusError
Error 130
Displayed if RF Gain could not be loaded. The
microprocessor on the lock controller board loads the D/A
converter in the Lock Receiver (IC10) via a serial Bus during
a Gain-setting. The microprocessor reads the digital Gain
value back in and compares it with the transmitted value. If
the values are not the same a Gain Bus Error is generated.
WR
~
CLK
CS_Gain
DATA
DATA_RETURN
LOCK HF Power Supply Failure
Error 157
All analog power supplies (LOCK_P15V, LOCK_N15V,
LOCK_P5V and LOCK_N5V) are checked by the
Microcontroller. The supplies are multiplied by a individual
factor and added to a common signal L-RX, IC18).
Possible Causes:
- One or several Supplies missing.
- Some Power Supplies do not fulfil their specifications.
MUX L-RX (IC19) Pin
10
LOCK_P15V
LOCK_N15V
LOCK_P5V
LOCK_5V
LOCK HWTesterIsOff
Error 50
Function is only available in Hardware Testmode.
LOCK HWTesterIsOn
Error 51
System is not available as the Hardware Testmode is active.
Switch to the normal Mode with ‘Reset’.
LOCK ModeBusError
Error 134
Appears if the Mode could not be loaded. The microprocessor
on the LCB loads the transmitter shift register (IC27) via a
serial Bus using a Mode change (e.g. FFA and Normal Lock
Mode). The mode settings are read into the microprocessor
again and compared with the transmitted values. If the values
differ, a Mode Bus Error is generated.
WR
~
CLK
CS_CNTR
DATA
STATUS_DATAR
LOCK No Communication with
DSP
Error 149
This error appears, if the DSP does not clear its host interface
receive registers. Normally, this registers are released after a
read cycle by the DSP semiconductor
Possible Causes:
- Protocol Failure at the Host Bus.
- Control Signals for Host Bus are not generated correctly.
LCB:
DSP_HEN~
DSP_HR_W~
uP: A0..A2
uP: D0..D7
Table 12. Lock Error Messages in alphabetical order
ERROR MESSAGE
ERROR DESCRIPTION
SIGNALS
Note: Signals of
the serial bus
(power, gain
phase, shift etc.)
are only active
during a process
change
Summary of Contents for DAEDALUS
Page 1: ...BRUKER BRUKER SPECTROSPIN AG SERVICE MANUAL DAEDALUS LOCK Version BSMS 004 ...
Page 6: ...Contents 6 Spectrospin AG Daedalus Lock Manual ...
Page 14: ...8 Spectrospin AG Daedalus Lock Manual ...
Page 16: ...10 Spectrospin AG Daedalus Lock Manual ...
Page 26: ...General Description 20 Spectrospin AG Daedalus Lock Manual ...
Page 36: ...Operation 30 Spectrospin AG Daedalus Lock Manual ...
Page 56: ...Lock Transmitter 50 Spectrospin AG Daedalus Lock Manual ...
Page 64: ...Lock Receiver 58 Spectrospin AG Daedalus Lock Manual ...
Page 68: ...19F RX Option 62 Spectrospin AG Daedalus Lock Manual ...
Page 85: ...Installation Daedalus Lock Manual Spectrospin AG 79 Installation 11 2 ...
Page 86: ...Z0 Compensation Option 80 Spectrospin AG Daedalus Lock Manual ...
Page 90: ...Technical Data 84 Spectrospin AG Daedalus Lock Manual ...
Page 92: ...Trouble Shooting 86 Spectrospin AG Daedalus Lock Manual ...
Page 98: ...BSMS service tool 92 Spectrospin AG Daedalus Lock Manual ...
Page 107: ...Error Descriptions Daedalus Lock Manual Spectrospin AG 101 ...
Page 108: ...Lock Error Messages 102 Spectrospin AG Daedalus Lock Manual ...
Page 114: ...Appendix 108 Spectrospin AG Daedalus Lock Manual ...
Page 122: ...116 Spectrospin AG Daedalus Lock Manual ...
Page 123: ...Daedalus Lock Manual Spectrospin AG 117 Lock Transmitter Index B ...
Page 124: ...118 Spectrospin AG Daedalus Lock Manual ...
Page 125: ...Daedalus Lock Manual Spectrospin AG 119 Lock Transmitter Index C ...
Page 126: ...120 Spectrospin AG Daedalus Lock Manual ...
Page 127: ...Daedalus Lock Manual Spectrospin AG 121 19F Transmitter Option Index A ...
Page 128: ...122 Spectrospin AG Daedalus Lock Manual ...
Page 129: ...Daedalus Lock Manual Spectrospin AG 123 Lock Receiver Index C ...
Page 130: ...124 Spectrospin AG Daedalus Lock Manual ...
Page 131: ...Daedalus Lock Manual Spectrospin AG 125 Lock Receiver Index D E ...
Page 132: ...126 Spectrospin AG Daedalus Lock Manual ...
Page 133: ...Daedalus Lock Manual Spectrospin AG 127 19F Receiver Option Index A ...
Page 134: ...128 Spectrospin AG Daedalus Lock Manual ...
Page 135: ...Daedalus Lock Manual Spectrospin AG 129 Lock Controller Index A ...
Page 136: ...130 Spectrospin AG Daedalus Lock Manual ...
Page 137: ...Daedalus Lock Manual Spectrospin AG 131 Lock Controller Index B C ...
Page 138: ...132 Spectrospin AG Daedalus Lock Manual ...
Page 139: ...Daedalus Lock Manual Spectrospin AG 133 Lock RS232 Biggy Board ...
Page 140: ...134 Spectrospin AG Daedalus Lock Manual ...
Page 141: ...Daedalus Lock Manual Spectrospin AG 135 Z0 Compensation Option ...
Page 142: ...136 Spectrospin AG Daedalus Lock Manual ...
Page 143: ...Daedalus Lock Manual Spectrospin AG 137 Cables ...
Page 144: ...138 Spectrospin AG Daedalus Lock Manual ...
Page 145: ...Daedalus Lock Manual Spectrospin AG 139 Test Equipment Index A ...