Utility Requirements
58
H157655_1_008
System and
Amplifiers
Mains Supply
Power
Consumpt
ion (kW)*
No. of
Spare
Electrical
Outlets
Length of
Mains
Cable (m)
TwoBay with
BL
BLAX300
230V 50/60 Hz / 16A single
phase; or 208V 60 Hz / 20A
single phase in USA
2.6
4 (230V
max. 10
amp.)
5.5
TwoBay with
BLAXH300/50
230V 50/60 Hz / 16A single
phase; or 208V 60 Hz / 20A
single phase in USA
2.2
4 (230V
max. 10
amp.)
5.5
TwoBay optimized
for solids 300 to
700 MHz
230V 50/60 Hz / 32A single
phase or 400V 50/60 Hz /
16A triple phase; or 208V 60
Hz / 30A single phase in
USA
5.0
4 (230V
max. 10
amp.)
5.5
OneBay with
BLAXH100/50
230V 50/60 Hz / 16A single
phase; or 208V 60 Hz / 20A
single phase in USA
1.6
2.2
4 (230V
max. 10
amp.)
5.5
NanoBay
230V 50/60 Hz / 16A single
phase; or 208V 60 Hz / 20A
single phase in USA
1.0
---
5.5
Imaging Cabinet
230V 50/60 Hz / 32A single
phase or 400V 50/60 Hz /
16A triple phase; or 208V 60
Hz / 30A single phase in
USA
2.4
---
---
Bayvoltex Chiller
for MicroImaging
Systems
230V 50/60 Hz / 16A single
phase or 208V 60 Hz / 20A
single phase in USA
approx.
0.45 kW
---
---
BCU-I/BCU-II
230V 50/60 Hz / 16A single
phase or 208V 60 Hz / 20A
single phase in USA (power
for BCU-II must come from
separate outlet).
0.45 kW
(BCU-I)
2.3 kW
(BCU-II)
---
---
UPS for CCU
UPS requirements: At least 5 kW for the AVANCE and CCU.
UPS for AVANCE
Spectrometer
Cabinet
UPS requirements: At least 2.6 kW for the spectrometer cabinet
(depends on configuration). The battery time must be selected
according to the maximum duration anticipated for a power failure.
* Includes the NMR workstation and computer monitor, and was measured using 2
amplifiers operating at maximum output in CW mode. For systems fitted with additional
amplifiers allow 300W for each additional amplifier.
Note: For the electrical specifications associated to the CryoProbe, CryoProbe - Prodigy,
and the Bruker Nitrogen Liquifier accessories, refer to the relevant pages in
.
Table 8.1:
Requirements of Basic System (2 Channels)
Summary of Contents for AVANCE
Page 6: ...Contents vi H157655_1_008 ...
Page 24: ...System Components 24 H157655_1_008 ...
Page 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Page 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Page 72: ...Utility Requirements 72 H157655_1_008 ...
Page 82: ...Floor Plan 82 H157655_1_008 ...
Page 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Page 102: ...Contact 102 H157655_1_008 ...
Page 104: ...List of Figures 104 H157655_1_008 ...
Page 109: ...H157655_1_008 109 ...
Page 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...