ZTKS0173 / Z31990 / 04
Technical Data MS 400‘54 Ascend ULH
_ _ _
ʹ
_ _ _
ʹ
_ _ _ _
9.8
Technical Data Magnet
Operating Modes of the Magnet System:
Driven Mode
In the driven mode the current lead is mounted and the current is led through the power
supply. The coils of the magnet can be energized and de-energized.
Persistent Mode
In the persistent mode the current flows exclusively within the magnet. The circuit has no
connection to the outside. The magnetic field cannot be switched off.
Technical Data Magnet
Value
Unit
Proton frequency
400
MHz
Central field
9.39
T
Coil inductance
30.39
H
Magnetic energy
164.5
kJ
Maximum drift rate
0.01
ppm/h
4
Hz/h
Table 9.11: Specification of the Magnet
This p
age
wi
ll b
e
rep
lace
d
by spe
cific
te
chn
ical
d
ata
.
Summary of Contents for Ascend 400'54
Page 1: ...Version Innovation with Integrity 400 54 Ascend ULH 04 User Manual NMR...
Page 6: ...6 ZTKS0173 Z31990 04 Table of Contents...
Page 8: ...8 ZTKS0173 Z31990 04 Contact...
Page 32: ...32 ZTKS0173 Z31990 04 Transportation...
Page 33: ...33 ZTKS0173 Z31990 04 4 Assembly 4 1 Safety Approved Persons Bruker Service only...
Page 34: ...34 ZTKS0173 Z31990 04 Assembly...
Page 55: ...55 ZTKS0173 Z31990 04 8 Disassembly 8 1 Safety Approved Persons Bruker Service only...
Page 56: ...56 ZTKS0173 Z31990 04 Disassembly...
Page 88: ...88 ZTKS0173 Z31990 04 Technical Data MS 400 54 Ascend ULH...
Page 90: ...90 ZTKS0173 Z31990 04 Warning Signs...
Page 92: ...92 ZTKS0173 Z31990 04 List of Figures...
Page 94: ...94 ZTKS0173 Z31990 04 List of Tables...
Page 96: ...96 ZTKS0173 Z31990 04 Glossary...
Page 98: ...Bruker Corporation info bruker com www bruker com Order No Z31990...