46
ZTKS0173 / Z31990 / 04
Troubleshooting
Table continued
Indicator
Possible reason
Solution
Performed by
Shim current can
not be set correctly.
The control cable is
not connected
correctly to the
current lead or to
the power supply.
Connect the control cable
correctly to the current lead
and to the power supply.
Bruker Service
The switch „Main
Coil/OFF/Shim
Coil“ is in the wrong
position.
Change the switch position.
Bruker Service
Shims do not affect
the NMR signal.
Shim heater
current is not
correct. The shim
switches are not
opened.
Set the shim heater current
to the specified value
(see
).
Bruker Service
Magnet system
does not reach
specification.
Magnetic material
inside RT bore
tube.
Carefully clean the RT bore
tube.
Bruker Service
Large ferromag-
netic parts near the
magnet system.
1.
As much as possible,
keep all ferromagnetic
parts away from the
magnet system.
2.
Repeat shimming.
Bruker Service
Summary of Contents for Ascend 400'54
Page 1: ...Version Innovation with Integrity 400 54 Ascend ULH 04 User Manual NMR...
Page 6: ...6 ZTKS0173 Z31990 04 Table of Contents...
Page 8: ...8 ZTKS0173 Z31990 04 Contact...
Page 32: ...32 ZTKS0173 Z31990 04 Transportation...
Page 33: ...33 ZTKS0173 Z31990 04 4 Assembly 4 1 Safety Approved Persons Bruker Service only...
Page 34: ...34 ZTKS0173 Z31990 04 Assembly...
Page 55: ...55 ZTKS0173 Z31990 04 8 Disassembly 8 1 Safety Approved Persons Bruker Service only...
Page 56: ...56 ZTKS0173 Z31990 04 Disassembly...
Page 88: ...88 ZTKS0173 Z31990 04 Technical Data MS 400 54 Ascend ULH...
Page 90: ...90 ZTKS0173 Z31990 04 Warning Signs...
Page 92: ...92 ZTKS0173 Z31990 04 List of Figures...
Page 94: ...94 ZTKS0173 Z31990 04 List of Tables...
Page 96: ...96 ZTKS0173 Z31990 04 Glossary...
Page 98: ...Bruker Corporation info bruker com www bruker com Order No Z31990...