28
&"
#"! !$! #
(98/12/10)
CCU DRAM configuration:
Select Jumper W6
SEL0 1 o––––o 2 SHORT
SEL1 3 o o 4 OPEN
00
1o––––o2
3o––––o4
16MB
2x8 MB
P/N 65173 or
P/N 67924
as U75, U76
10
1o o2
3o––––o4
32MB
2x16 MB
P/N 66123 or
P/N67946
as U75, U76
11
1o o2
3o o4
64MB
2*32 MB
P/N 66124 or
P/N67945
as U75, U76
Note: Any installed DRAM module is not available if jumper W7 is removed
The new CCU version need following software
Boot PROM version:
970611
Diskless version:
970501 included XWIN NMR 2.0 re-
lease
3. 8. Introduction Status of the AQS CCU10
3. 8. 1. Modifications of the introduced layout
The new AQS CCU CU10 board contains the same functions as the previous CU09 P/N
H2570
excepted the middle J0/ST2 connector for the backplane which has been changed . Addi-
tionally, the RS485 interface is connected to the J0/ST2 connector and now available on
the backplane X3 connector. Therefore, it is physically not compatible with the old AQX
system.
3. 8. 2. Jumper Configuration
Normal setting of installed jumpers:
W4
A+B (16 Mbyte DRAM with 2 8Mbyte SIMM’s)
W1
not set
W12
C
W6
not set
W10
set
W11
not set
Summary of Contents for AQS
Page 1: ...BRUKER Version Communication Control Unit CCU 9 10 AQS AQX Technical Manual 001...
Page 3: ...AQX CCU 9_4x00 AQS CCU 10_4x00...
Page 23: ...W5 W4 W3 Test jumper W6 DRAM DRAM W2 W1 W7 High Low Firmware PROM s U75 U76...
Page 24: ...W1 High Low Firmware PROM s W7 U75 U76 W5 W4 W3 Test jumper W6 DRAM DRAM W2...
Page 26: ...CU10 ST8 ST9 tty01 tty09 RS232 Connector CU10 ST2 tty10 tty20 RS485 JTAG and BBIS Connector...