OFF indicates that the PHY did not find a valid link on the net-
work connection. Transmit and receive operation are not pos-
sible. 
In this situation the hardware is performing Autonegotiation
without any software interaction.

”Receive Activity”

ON indicates that a packet is being received in 10BASE–T or
100BASE–TX mode.

”Transmit Activity”

ON indicates that a packet is being transmitted in 10BASE–T
or 100BASE–TX mode.

2. 9.  Operational Settings

2. 9. 1. NVRAM Setting

The NVRAM contains a set of parameters which are used by the software. After power
up, the monitor checks the parameter called 

magic

.. If 

magic

. is anything else than

RISCPROM

, e.g. if a new and empty NVRAM has just been inserted, the monitor does the

whole set of diagnostic tests and sets the NVRAM parameters to the following default
values:

netaddr=0x149.236.99.99

netmask=0xffffff00

lbaut=9600

rbaut=9600

bootfile=bfs()/usr/diskless/clients/spect/root/unix

bootmode=c

console=0

cpuid=0

resetpc=xxxxxxxx

resetra=xxxxxxxx

eth_addr_ch1=0x

eth_addr_ch2=0x

version=5.40

magic=RISCPROM

model=CCU R4600

vendor=BRUKER

rootname=0

use_bootparams=1

keyswtch=0

keyboard=AT

This procedure takes a few minutes. The monitor prints a test protocol and at last its
prompt (

>>

) on 

tty0

 which is the default 

console output

.

The NVRAM parameters can be printed with the monitor by

printenv

and can be changed by

setenv 

parameter value

A problem crops up if 

magic

 is equal to 

RISCPROM

 and 

lbaud

 and 

rbaud

 doesn’t meet

the bauderate of the connected terminal or 

console

 doesn’t meet the interface to which

the terminal is connected to. Then no printout or input is possible. To correct this

Summary of Contents for AQS

Page 1: ...BRUKER Version Communication Control Unit CCU 9 10 AQS AQX Technical Manual 001...

Page 2: ...the manual leading to coincidental damage whether during installation or operation of the instrument Un authorised reproduction of manual contents without written permission from the publishers or tra...

Page 3: ...AQX CCU 9_4x00 AQS CCU 10_4x00...

Page 4: ...s 19 2 11 Power Requirements 22 3 Service Information 23 3 1 Operating Conditions 23 3 2 Network connection 23 3 3 Software Implementation 24 3 4 Problems of Point to Point Connection to the O2 25 3 5...

Page 5: ...igure 1 CU09 Controls at Front Panel 6 Figure 2 CU10 Controls at Front Panel 7 Figure 3 CU09 Attaching SubD connectors to the RS232 485 interfaces 10 Figure 4 CU 10 Attaching SubD connectors to the RS...

Page 6: ...ssories 8 Table 6 CU09 Configuration of Serial Interfaces 10 Table 7 CU10 Configuration of Serial Interfaces 11 Table 8 Parameters to be set 15 Table 9 DRAM configuration 16 Table 10 RS232 Signal assi...

Page 7: ...are connected inside to the same signals but have crossed pin assignments The upper socket signed FETH HUB has to be used if CU09 10 will be connected to an HUB This is the normal way with twisted pa...

Page 8: ...up depending on the interface to which the spect is connected to 1 4 Problems of Point to Point Connection to the O2 Both link partners of a Fast Ethernet Connection exchange their abilities 10 or 100...

Page 9: ...ns Type on O2 as superuser vi etc config netif options This file contains the 2 parameter entries of if1name ec0 and if2name ec1 Swap the values of both parameters to if1name ec1 if2name ec0 Type wq i...

Page 10: ...memory to IO data path is 32 bit wide and runs at 25 MHz AQX Software Constrains CCU Versions Part No Layout No EC Level AspectStation SGI Computers 2 2 Former AQS CCU Versions The AQS_CCU is a CPU b...

Page 11: ...RS485 channels using Z85C230 labeled as tty10 and tty20 connectable through one 8 pin Mini D console and two 50 pin con nectors tty01 tty09 at the front edge real time clock MK48T02 with 2 kbyte non v...

Page 12: ...Wire BNC connector Reset Button RS232 console Mini D 8 Activity LED s TTY01 TTY05 TTY10 TTY06 TTY09 TTY20 10 100 BASE T Fast Ethernet Status LED s to Hub s Fast Ethernet connector RJ 45 to Host VME ac...

Page 13: ...00 BASE T Fast Ethernet Status LED s to Hub s Fast Ethernet connector RJ 45 to Host VME access Memory access 10Base2 Ethernet activity Collision occurence Good Link status Receive activity Transmit ac...

Page 14: ...U09 Fast Ethernet UTP CAT5 RJ45 Cable 5m 83025 CU09 Fast Ethernet UTP CAT5 RJ45 Cable 10m 83026 CU09 Fast Ethernet UTP CAT5 RJ45 Cable 0 5m 67980 CU09 Fast Ethernet UTP CAT5 RJ45 Cable 3m 67981 CU09 F...

Page 15: ...ne or two cables of the following type SCSI Cable P N 73104 Connecting to RS232 RS485 of CU10 The RS485 Channels tty10 and tty20 are only avalable at the AQS Backplane Connec tor X3 Connecting the RS2...

Page 16: ...6 tty07 tty08 tty09 tty01 tty02 tty03 tty04 tty05 not used not used CAN tty06 tty07 tty08 tty09 Connections made with SCSI Cable P N 73104 Thin wire Ethernet BNC Connector RJ45 Connector to other Host...

Page 17: ...tty08 tty09 tty01 tty02 tty03 tty04 tty05 not used not used CAN tty06 tty07 tty08 tty09 Connections made with SCSI Cable P N 73104 Thin wire Ethernet BNC Connector RJ45 Connector to other Hosts St5 RJ...

Page 18: ...c tion CU09 Terminal console cable 3m P N HZ04112 This cable was originally designed to connect a plotter to the Indy Figure 5 Console Cable MINI D 8 pin to D_SUB 9 pin Pin Signal 1 DTR 2 CTS 3 TxD 4...

Page 19: ...flect the activities of the Ethernet DMA the CPU access to DRAM and the activity of the CPU on VME bus The intention is that after power up or reset a basic test procedure changes the LED value at suc...

Page 20: ...le set of diagnostic tests and sets the NVRAM parameters to the following default values netaddr 0x149 236 99 99 netmask 0xffffff00 lbaut 9600 rbaut 9600 bootfile bfs usr diskless clients spect root u...

Page 21: ...or clears the memory after reset and boots without doing any test d The boot procedure has to be started by typing u or auto the memory will not be cleared and no tests will be carried out eth_addr_ch...

Page 22: ...sition A disables the SQE test mode e g if there is a Repeater connected 2 9 3 Installing DRAM It is possible to equip the CCU with 16 32 or 64 Mbyte using 2 of 8 Mbyte DRAM modules P N 65173 16 Mbyte...

Page 23: ...W5 W4 W3 Test jumper W6 DRAM DRAM W2 W1 W7 High Low Firmware PROM s U75 U76...

Page 24: ...W1 High Low Firmware PROM s W7 U75 U76 W5 W4 W3 Test jumper W6 DRAM DRAM W2...

Page 25: ...ons CU09 10 INDY O2 Terminal Mini DIN 8 Mini DIN 8 D SUB 9 D SUB 25 Shield housing housing housing 1 DTR 1 5 CTS 2 6 7 20 TxD 3 5 2 3 GND 4 4 5 7 RxD 5 3 3 2 RTS 6 2 8 6 DCD 7 4 GND 8 CU09 ST8 ST9 tty...

Page 26: ...CU10 ST8 ST9 tty01 tty09 RS232 Connector CU10 ST2 tty10 tty20 RS485 JTAG and BBIS Connector...

Page 27: ...standard twisted pair cable has a straight through pin to pin connection from plug to plug The star topology of twisted pair Ethernet connecting several Hosts via one Hub requires a different signal...

Page 28: ...oltages Part No 5 V 12 V 12 V 5 V analog J3 C8 5 V analog J3 C1 C5 CCU CU09 10 H2570 H9503 5 3 A 0 3 A 0 1 A 0 0 CU09 supplies fused current at the following connectors ST8 9 RS485 12 Volt 0 2 Ampere...

Page 29: ...can be carried out as on CU09 with the exception of the RS485 interface that is now available on the AQS backplane The BNC cable of the Thin Wire Ethernet 10Base2 should be left unconnected 10Base2 is...

Page 30: ...ne of the two link partners is down by software or power The Fast Ethernet cable can be pulled out and plugged in under power and activity Autonegotiation and link checking starts on both partners wit...

Page 31: ...reases the performance by at least 10 and leads to collisions Furthermore the point to point connection with O2 causes errors during the boot se quence This problem can be solved by connecting the O2...

Page 32: ...r diskless clients spect root unix This call leads to booting over 10Base2 The CU09 10 monitor will boot over Fast Ether net if fe is inserted instead of la or as default if no name is set boot f bfs...

Page 33: ...OUI Device Group Series Production 0000ad 0000 09 3 7 Introduction Status of the AQX CU09 CCU 3 7 1 Modifications of the introduced layout There are 25 prototypes of CU09 using layout H3P2160C with s...

Page 34: ...on 970501 included XWIN NMR 2 0 re lease 3 8 Introduction Status of the AQS CCU10 3 8 1 Modifications of the introduced layout The new AQS CCU CU10 board contains the same functions as the previous CU...

Page 35: ...5 CCU Connector ST7 ST8 ST9 ST9 ST2 ST2 AQX Ext U Connector N A ST2 ST 6 ST7 ST 8 ST9 ST10 N A N A Pin out of the X3 connector at the AQS backplane Signal Signal 2 TxD _1 1 TxD _1 4 GND 3 GND 6 RxD _1...

Page 36: ...to be virgin Starting with Firmware Version 199990408 the Power on Diagnostics will be first carried out at the second boot procedure If magic RISCPROM the Firmware will boot normally but if the para...

Page 37: ...th_addr_ch1 0x0000ad03xx06 eth_addr_ch2 0x0000adxxxx09 version 5 40 magic RISCPROM model CCU R4600 vendor BRUKER rootname 0 use_bootparams 1 keyswtch 0 keyboard AT 3 8 3 2 Firmware CU09 10 needs the f...

Page 38: ...rate 15 to 255 Reserved and zero The serial boot prom is listed in the PAL part list It is named CU09EA00 and labeled EA or 50MHz 3 9 AQX CU09 CCU History of Modifications EC No Date Part Number Descr...

Page 39: ...can access only the Memory 3 Slave devices which are accessible only by the CPU The communication between these Sections is provided by Interconnect Interface This is a CCU s Internal Interface resolv...

Page 40: ...Data reg Triport bus Exchanger registered P L M rocessor Bus Bus emory Bus Ae L Address Bus L Data Bus P Bus Pipeline r Memory Write multil reg 4 x 64 bit Memory 64 64 Bit Wide 32 9 De 32 M_D Mma P_D...

Page 41: ...e referred to as word 32 bit operands with the valid bytes indi cated by b at Byte format see below Byte 0 means the byte connected to data lines D0 to D7 Byte halfword 2 byte word 4 byte and block 4...

Page 42: ...This RAM has to be en abled by setting jumper W3 and is then located in the lowest address space starting with address 0x0 The installed memory volume must be properly selected by setting of jumper W...

Page 43: ...e Pagemode is not recommended because of its impact to the on board ethernet operation Single word writes and reads use a single read write register A pipeline structure is made for multiple writes an...

Page 44: ...eserved R W bbbb 1F08003C CFIT Interrupt R W bbbb undefined 1F080040 CFDA Driver area R W bbbb undefined Control and Status Register Device Codes The CSRs registers are accessibly only after loading a...

Page 45: ...and Fiber Optic cables The chip interfaces the PMD sublayer through DP82223 Twisted Pair Transceiver to the MAC layer 21140A through a Media Independent Interface MII The 32 word wide MII status and...

Page 46: ...to 0001 05h ANLPAR 5 Does not reflect Link Partner 10BASE T Abil ity after parallel detection Reflects Link Partner 10BASE T Ability after parallel detection 05h ANLPAR 7 Does not reflect Link Partner...

Page 47: ...6 21 1 6 5 762 26 6 21 5 1 19h 11 5 48 55 46 55 46 6 4 5 4 0 4 6 0 276 Differences between DP83840 and 83840A in full duplex mode 00h 9 9 01h 9 9 02h 9 9 03h 9 9 8 5 21 2 04h 9 9 05h 9 9 1 461 4 06h...

Page 48: ...F022xxx Channel A Control Register cons R W x x x b 1F023xxx Channel A Data Register cons R W x x x b 1F024xxx Channel D Control Register tty03 R W x x x b 1F025xxx Channel D Data Register tty03 R W x...

Page 49: ...to 1F0A0000 1F0A3FFF address region JTAG submodul is compatible with IEEE 1149 1 JTAG Standard Test Access Port and Boundary Scan Architecture Supported National s SCAN Ease Embedded Application Softw...

Page 50: ...rmally protected and can only be writ ten by means of the special PAL device CU09Nx01 of the service staff So far this range contains only the ethernet physical address Besides the timing registers th...

Page 51: ...s not available if the respective digit equals to zero All configuration register outputs are provided in their negated form The STATUS register accommodates the temperature and VME bus time out statu...

Page 52: ...ssigned and hardwired to the floating point in terrupt The router distributes all onboard and vme bus interrupt sources to level 0 to 4 Source interrupts intended to a certain level are ored to produc...

Page 53: ...d their location in the corresponding interrupt regis ter are as follows Interrupt Level Location INT_0 INT_1 INT_2 INT_3 INT_4 in Reg IRQ1 IRQ4 IRQ6 0 IRQ7 D0 TMP_INT IRQ3 IRQ5 0 F_INT D1 res TIM_INT...

Page 54: ...implemented W x x x b 4 2 12 CIS access The information EEPROM of the Component Identification System is connected to the CCU con trolled CIS channel The EEPROM of all CCU s have one and the same uni...

Reviews: