BN1100 Revision: 0
Initial Issue: 21/08/2012
ATA 31-10-04
Fig. 5
Mounting the transducer on the calibrator:
3)
Start the calibrator by pressing its button. The level displayed / recorded by the DMAU must be:
•
Type 4511-W-006 –
Monoaxial
:
10 mV (accelerometer sensitivity × 10 ms
2
× gain of DMAU) ± 8%
•
Type 4524-W-001 –
Triaxial
:
100 mV (accelerometer sensitivity × 10 ms
2
× gain of DMAU) ± 8%
Note:
With Type 4524-W-001
–
Triaxial
, , the three positions on the WA-1732 allow to test of the three axes X, Y and Z.: