Operational Matrix for the Model DO600 

Function / 

Resulting Action 

Power

Status 

Mode 

Setting 

Required Button Press Sequence 

Comments 

On/Off 

Any 

Any 

Momentary press of the ON/OFF button 

 

Water Saturated 

Air Calibration 

On 

Any 

Place Electrode in Calibration Cap 

Press & hold CAL/RECALL button for 2 seconds 

Zero Calibration 

On Any 

Place Electrode in Zero Solution, wait for stability 

Press & hold CAL/RECALL button for 2 seconds 

Also works with probe 

removed

(Cal at zero current) 

Store Reading 

On 

Any 

Momentary press of the MODE/HOLD button 

Stores and holds 

reading “HOLD” 

displayed 

Hold Release 

On 

While In 

Hold Mode 

Momentary press of the MODE/HOLD button 

 

Enter Memory 

Retrieval 

On Any 

Momentary press of the Cal button followed by a 

momentary press of the MODE/HOLD button (w/in 4 

seconds)

If no data is stored in 

memory, “End” is 

displayed briefly, and 

then returns to last 

mode. 

Scroll Stored 

Readings 

On 

Memory

Recall 

Momentary press of the MODE/HOLD button 

 

Exit Memory 

Retrieval 

On 

Memory

Recall 

Momentary press of the CAL/RECALL 

 

Clear Stored 

Memory On 

Any

Measure-

ment Mode 

Press and hold the ON/OFF button for 4 seconds 

“clr” is displayed. 

Change

Measurement 

Mode

On Any 

Press and hold the MODE/HOLD button for at least 

two seconds (the modes will scroll until the button is 

released) 

Enter Salinity 

Compensation 

On Any 

Press and release the CAL/RECALL button twice 

(momentary) in succession (Displays SAL) 

Changing Salinity 

Compensation On  SAL 

Momentary press of the MODE/HOLD button (each 

button press increases the ratio by 1 ppt (part per 

thousand), the value cycles from 0 to 50) 

Exit Salinity 

Compensation 

On SAL 

Press and release CAL/RECALL button for 2 seconds 

to enter Altitude Compensation, Or 

Press and release Cal Button one more time to enter 

measurement mode 

It is necessary to 

press the 

CAL/RECALL button 

to save changes. If 

the unit times out, no 

changes are saved. 

Enter Altitude 

Compensation 

On 

Any or SAL 

Press CAL/RECALL twice (momentary). Unit enters 

Salinity Mode. Press CAL/RECALL for 2 seconds to 

enter Altitude Compensation Mode. (Displays Ald) 

Unit times out in 5 

seconds if no button 

is pressed, reverts to 

previous mode. 

Changing Altitude 

Compensation On  Ald 

Momentary press of the MODE/HOLD button(each 

button-press increases altitude by 1,000ft, the value 

cycles from 0 to 20) 

Exit Altitude 

Compensation 

On Ald 

Momentary press CAL/RECALL button to exit and 

save changes. 

It is necessary to 

press the 

CAL/RECALL button 

to save changes. If 

the unit times out, no 

changes are saved. 

Change

Temperature 

Units

Off 

n/a (off 

mode)

Press and hold the CAL/RECALL button then 

momentarily press the ON/OFF button.  Release the 

CAL/RECALL button after the "SELF CAL" lights 

Override 

Autopower Off 

On 

Any 

Press CAL/RECALL button (momentary) then 

simultaneously press and hold the MODE/HOLD and 

ON/OFF buttons for for 2 seconds. 

Default Reset 

Off 

n/a (off 

mode)

Simultaneously press ON/OFF, CAL/RECALL and 

MODE/HOLD momentarily. “dFLt” will be displayed. 

Summary of Contents for Vinoferm DO-600

Page 1: ...eling automatische temperatuurcompensatie zoutgehalte en hoogtecompensatie Als u deze meter voorzichtig gebruikt en onderhoudt zal hij u jarenlang trouwe diensten bewijzen Beschrijving van de meter Beschrijving frontpaneel 1 Deksel batterijvak 2 LCD display 3 MODE HOLD toets 4 CAL RECALL toets 5 AAN UIT toets 6 Elektrodekraag 7 Opgeloste zuurstofsensor 8 Borgkapconstructie voor het membraan 9 Memb...

Page 2: ...was last turned OFF will remain on the display 3 If this is the first time the meter is powered up wait ap prox 3 minutes for the probe to polarize please disable auto power off Instructions for disabling auto power off are provided in a later section For more details please read the Startup Polarization Period paragraph earlier in this manual before continuing 4 Select the desired units of measur...

Page 3: ...es 20 000ft above sea level 4 Momentarily press the CAL RECALL button to save the compensation setting and return to the normal mea surement mode Auto Power OFF Feature The auto power off feature automatically shuts the meter off 10 minutes after the most recent button press To disable this feature refer to the Disable Auto Power OFF section Disable Auto Power OFF Feature With the unit ON momentar...

Page 4: ...it disconnects 3 To attach an electrode align the positioning keys on the electrode and the main body housing and then care fully push the electrode into the meter socket until it is fully seated 4 Tighten the electrode retaining collar firmly enough to seal the electrode with the meter DO Bonded Membrane Cap replacement Important Note Do not touch the membrane as skin oils will interfere with the...

Page 5: ...ne Bonded membrane cap with threaded fitting Operating Temp Range 0 to 50 C 32 to 122 F ATC Range 0 to 50 C 32 to 122 F Salinity Compensation 0 to 50 ppt in 1 ppt part per thousand increments Altitude compensation 0 to 20 000 ft in 1000 ft increments Measurement Storage 25 tagged numbered data sets with recall Battery Power Four 4 CR2032 button batteries Low Battery Indication BAT appears on the L...

Page 6: ...is depleted Replace electrolyte and membrane cap assembly Readings drift down Insufficient stirring the probe consumes oxygen at the measuring surface requiring constant sample movement across membrane Move probe in sample or stir sample Slow Response Dirty or damaged membrane Replace electrolyte and membrane cap Electrode cannot be calibrated Depleted electrolyte Dirty or damaged membrane Replace...

Page 7: ...nity Compensation On Any Press and release the CAL RECALL button twice momentary in succession Displays SAL Changing Salinity Compensation On SAL Momentary press of the MODE HOLD button each button press increases the ratio by 1 ppt part per thousand the value cycles from 0 to 50 Exit Salinity Compensation On SAL Press and release CAL RECALL button for 2 seconds to enter Altitude Compensation Or P...

Page 8: ...ELF CAL lights Override Autopower Off On Any Press CAL RECALL button momentary then simultaneously press and hold the MODE HOLD and ON OFF buttons for for 2 seconds Default Reset Off n a off mode Simultaneously press ON OFF CAL RECALL and MODE HOLD momentarily dFLt will be displayed Re order and Accessory Information Part Number Description DO600 ExStik II Dissolved Oxygen Meter DO605 Replacement ...

Reviews: