The refillable electrode does not need to be detached from the body of the
ExStikTM in order to perform the refilling procedure.
Refillable electrodes (PH115) have a removable reference junction (slotted) and
the word REFILLABLE on the side of the electrode housing.
Removing the Reference Junction
The removal tool supplied with the PH113 Refill Kit is used to remove
the reference junction from the sensing surface of the electrode. If the reference
junction does not have slots for the ‘teeth’ of the removal tool to lock onto, the
electrode is NOT refillable.
Holding the electrode upside down, unscrew and remove the reference junction
using the removal tool.
Filling the Electrode
1. Once the reference junction is removed, fill the cavity with the refill solution
supplied in the Refill Kit.
2. Replace the reference junction using the removal/installation tool. (Spare
junctions
are provided if needed).
Filling Solution
The supplied container includes 15ml of filling solution. There is enough solution for 4 to 5 refills.
Use only the supplied solution for refilling the electrode.
Brouwland
Korspelsesteenweg 86 • B-3581 Beverlo - Belgium
Tel. +32-(0)11-40.14.08 • Fax. +32-(0)11-34.73.59
[email protected] • www.brouwland.com
Manual
013.130.0 pH meter pH110
5 / 6