Data Hold
Momentarily press the MODE button to freeze the current reading. The HOLD display icon will appear along with
the held reading. The held reading will also be stored in memory.
Momentarily press the MODE key to return to normal operation.
15-Storing Readings into Memory
1. Momentarily press the MODE button to store a reading. The LCD will briefly display the memory location
number and then the value stored (Data Hold will activate).
2. Momentarily press MODE again to return to normal operation.
3. Repeat step 1 to store the next reading and so on.
4. After 15 readings are stored the ExStikTM will return to memory location 1 and start overwriting existing data
with newly stored data.
Recalling Stored Readings
Note: Check that the HOLD symbol is not displayed. If it is, exit the HOLD function by momentarily pressing the
MODE button.
1. Momentarily press the CAL button and then press the MODE button immediately after CAL is displayed; the
storage location number (1 through 15) will flash. If the CAL mode is accidentally accessed (display flashing),
press the CAL button again to exit.
2. The last reading stored will be displayed first. To advance through the stored readings, momentarily press the
MODE button. The location number is displayed first, followed by the reading stored in that location.
3. To exit the recall mode, momentarily press the CAL button and the ExStikTM will return to normal operation.
CAL Reminder Display
When the ExStikTM is turned on in the pH mode for the 15th time without recalibration, the ‘CAL’ icon appears on
the LCD indicating that the ExStikTM may require calibration. Some applications may require recalibration of the
electrode more frequently than others. The CAL display is simply a reminder and will turn off when the pH
electrode is recalibrated.
RENEW Display
A flashing ‘RENEW’ warning indicates that the probe may be nearing the end of its useful life. If cleaning or
recalibration does not cause the RENEW icon to disappear, replace the electrode. The RENEW display appears
when the output of the pH electrode fails a diagnostic test.
Considerations
• If the unit appears to be locked (display frozen) it is possible that the Data Hold mode has been inadvertently
accessed by pressing the MODE button. Simply press the MODE button again or turn the meter off and restart if
the display appears frozen.
• If the meter does latch up and no button presses revive it, remove the batteries, push the ON button for 3
seconds and then reinsert the batteries.
• Note that if the batteries are removed, any stored readings will be discarded. Also, the user calibration data for
pH will be cleared. New user pH calibration data is required. Factory calibration data for all models will be retained,
however.
Battery replacement
1. Twist off the battery compartment cap
2. Replace the four (4) 2032 batteries observing polarity.
3. Replace the battery compartment cap
Brouwland
Korspelsesteenweg 86 • B-3581 Beverlo - Belgium
Tel. +32-(0)11-40.14.08 • Fax. +32-(0)11-34.73.59
[email protected] • www.brouwland.com
Manual
013.130.0 pH meter pH110
4 / 6