background image

 

 

 

pH meter pH110 

 

Description  

 

Front Panel Controls 

1. Battery compartment cap 
2. LCD Display 
3. MODE / HOLD button 
4. CAL / RECALL button 
5. ON/OFF button 
6. Electrode collar 
7. Electrode (refillable on PH110) 
(Electrode cap is not shown)

 

Display 

1. Bargraph reading 
2. Measurement reading 
3. BAT (low battery) and 
HOLD (data hold) indicators 
4. Temperature display 
5. Bargraph scale designations 
6. Units of measure 
7. Calibration indicators 
8. RENEW and CAL indicators

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

WARNINGS 

- This appliance is not a toy and must not reach the hands of children. It contains hazardous objects as 
well as small parts that children could swallow. In case a child swallows a piece, please contact a 
physician immediately. 
- Do not leave batteries and packing material lying around unattended, they can be dangerous for 
children if they use them as toys. 
- If the appliance is not used for a long period, remove the batteries to prevent leaking. 
Used or damaged batteries can cause cauterization on contact with the skin. Therefore, always use 
gloves suitable for this occasion. 
- Check that the batteries have not been short-circuited. Do not dispose of batteries in fire. 
 
 
 
 
 
 

 

Brouwland  

Korspelsesteenweg 86 • B-3581 Beverlo - Belgium 
Tel. +32-(0)11-40.14.08 • Fax. +32-(0)11-34.73.59 
[email protected] • www.brouwland.com

 

Manual 

 

 

013.130.0 pH meter pH110    

1 / 6 

Summary of Contents for pH110

Page 1: ...us objects as well as small parts that children could swallow In case a child swallows a piece please contact a physician immediately Do not leave batteries and packing material lying around unattended they can be dangerous for children if they use them as toys If the appliance is not used for a long period remove the batteries to prevent leaking Used or damaged batteries can cause cauterization o...

Page 2: ...d acidic solutions with a pH higher than 7 are known as bases and solutions with a pH of exactly 7 are neutral The pH scale is logarithmic so for example if sample A is 1 pH less than Sample B this means that Sample A is 10 times more acidic then Sample B A difference of 1 pH represents a ten fold difference in acidity Getting Started Remove the cap from the bottom of the ExStikTM to expose the el...

Page 3: ...r 3 points A two point calibration with a buffer of 7 plus 4 or 10 whichever is nearest to the expected sample value is always recommended A one point calibration choose the value closest to the expected sample value is also valid For best accuracy always calibrate at the sample temperature 1 Place the electrode into a buffer solution 4 7 or 10 and momentarily press the CAL key pH 7 should be cali...

Page 4: ...Display When the ExStikTM is turned on in the pH mode for the 15th time without recalibration the CAL icon appears on the LCD indicating that the ExStikTM may require calibration Some applications may require recalibration of the electrode more frequently than others The CAL display is simply a reminder and will turn off when the pH electrode is recalibrated RENEW Display A flashing RENEW warning ...

Page 5: ...to the electrode is NOT refillable Holding the electrode upside down unscrew and remove the reference junction using the removal tool Filling the Electrode 1 Once the reference junction is removed fill the cavity with the refill solution supplied in the Refill Kit 2 Replace the reference junction using the removal installation tool Spare junctions are provided if needed Filling Solution The suppli...

Page 6: ...ter long storage without use rinse with demineralized water and leave to steep one day in preserving fluid If the measured value appears slowly this indicates reduced electrode activity If you have a refillable electrode that you wish to refill first bleed the electrode The electrolyte is no longer fit for purpose and will have to be replaced So always replace all of the electrolyte After refillin...

Reviews: