MASS-STREAM
TM
Instruction Manual D-6300
9.17.104B
– Date of issue: 2018-12
Page 28 off 34
Readout of bus-address/MAC-ID and baud rate:
Pressing the micro switch 3 times briefly with intervals of max 1 second in normal
running/operation mode will trigger the instrument to “show” its bus address/MAC-ID and
baud rate. For the indication of the bus-address/MAC-ID the green LED will flash the
amount of tens and the red LED the amount of units in the number. For the indication of
the baud
rate setting, both LED’s will flash. The flashes are called “count-flashes” and
have a pattern of 0.5 sec on / 0.5 sec off.
Table 5:
LED indications for bus address/MAC-ID and baud rate
green LED
red LED
Time
Indication
amount of count
flashes (0…12)
off
0…12 sec
maximum
tens in bus address/MAC-ID
for instrument
off
amount of count
flashes (0…9)
0…9 sec
maximum
units in bus address/MAC-ID
for instrument
amount of count
flashes (0…10)
amount of count
flashes (0…10)
0…10 sec
maximum
baud rate setting for instrument
Please note:
The value
’zero’ will be indicated by a period of
1 sec off (0.5 sec off / 0.5 sec off)
Examples:
For bus address /MAC-ID 35
the green LED will flash 3 times and the red LED will flash 5 times
For bus address /MAC-ID 20
the green LED will flash 2 times and the red LED will flash 0 times
For bus address /MAC-ID 3
the green LED will flash 0 times and the red LED will flash 3 times
For bus address /MAC-ID 126
the green LED will flash 12 times and the red LED will flash 6 times