MASS-STREAM
TM
Instruction Manual D-6300
9.17.104B
– Date of issue: 2018-12
Page 27 off 34
Table 4:
LED indications using micro switch
at power-up situation of an instrument
green LED red LED
Time
Indication
off
off
0
– 4 sec
No action
Pressing a switch shortly by accident will not
cause unwanted reactions of the instrument
off
normal
wink
4
– 8 sec
Restore parameters
All parameter settings (except fieldbus settings)
will be restored to situation of final test at
production
normal
flash
off
8
– 12 sec
FLOW-BUS
PROFIBUS DP /
Modbus-RTU
DeviceNet
TM
Auto install on
bus:
Instrument will
install itself to a
(new) free node
address
Not used
Not used
normal
wink
normal
wink
12
– 16 sec
Remote/manual installation
Instrument will get a default address. This can
be changed by means of micro-switch and
LED’s. See description below for description of
this procedure.
Specific fieldbus for default installation address:
FLOW-BUS
PROFIBUS DP /
Modbus-RTU
DeviceNet
TM
Node address = 0
Station address
= 126
MAC-ID = 63
Please note:
normal flash = 0.2 sec on / 0.2 sec off
Set instrument to default installation address/MAC-ID:
At first the instrument needs to be forced into the
“remote/manual install mode”. During
power-up the micro switch needs to be pressed-
down and released when both LED’s
indicate
‘normal flash’ (pattern: 0.2 sec on / 0.2 sec off). For more details see also
table 4
.
When the switch is released, instrument will get the
“default installation address”. This will
be different for each fieldbus system. The instrument may be switched-off now or it will
return into the
“normal running/operation mode” automatically after 60 sec. The baud rate
setting will not be changed by this action.