BRONKHORST HIGH-TECH B.V.
page 10
9.17.044
2 INSTALLATION
2.1 Receipt of equipment
Check the outside packing box for damage incurred during shipment. Should the packing box be damaged,
then the local carrier must be notified at once regarding his liability, if so required. At the same time a report
should be submitted to:
BRONKHORST HIGH-TECH B.V.
RUURLO HOLLAND
If applicable, otherwise contact your distributor.
Remove the envelope containing the packing list; carefully remove the equipment from the packing box.
Do not discard spare or replacement parts with the packing material and inspect the contents for damaged or
missing parts.
2.2 Return
shipment
When returning material, always describe the problem and if possible the work to be done, in a covering
letter.
It is absolutely required to notify the factory if toxic or dangerous fluids have been metered with the
instrument!
This to enable the factory to take sufficient precautionary measures to safe-guard the staff in their repair
department. Take proper care of packaging. If possible use the original packing box. Seal instrument in
plastic etc.
Contaminated instruments must be dispatched with a completely filled in 'declaration on
contamination form'.
Contaminated instruments without this declaration will not be accepted.
Note:
If the instruments have been used with toxic or dangerous fluids the customer should pre-clean the
instrument.
Important:
Clearly note, on top of the package, the customer clearance number of Bronkhorst High-Tech B.V., namely:
NL801989978B01
If applicable, otherwise contact your distributor for local arrangements.
2.3 Service
If the equipment is not properly serviced, serious personal injury and/or damage to the equipment could be
the result. It is therefore important that servicing is performed by trained and qualified service personnel.
Bronkhorst High-Tech B.V. has a trained staff of servicemen available.
2.4 Mounting
Install the LIQUI-FLOW in accordance to the direction of the FLOW arrow. The arrow for flow direction is
indicated on the instrument, between process fittings.
2.5 Fluid/gas
connections
Bronkhorst High-Tech B.V. LIQUI-FLOW meters/controllers are equipped with compression or face-seal-
fittings. For LIQUI-FLOW these fittings are orbitally welded to the body. For leak tight installation of
compression type fittings be sure that the tube is inserted to the shoulder in the fitting body and that no dirt or
dust is present on tube, ferrules or fittings. Tighten the nut finger-tight; while holding the instrument, then
tighten the nut 1 turn.
If applicable follow the guidelines of the supplier of the fittings. Special types of fittings are available on
request.
* Note:
Always check your system for leaks, before applying fluid/gas pressure. Especially if toxic, explosive
or other dangerous fluids are used.