Bus Address
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…255
199
125/10
0x0FAA/4011
Baud Rate
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned long
RW
0…1.0E10
201
125/9
0xFD48…0xFD49/64841…64842
Bus 1 Parity
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…2
335
125/12
0x0FAC/4013
The following values are supported:
Value
Description
0
No parity
1
Odd parity
2
Even parity
Communication via standard connection (RS232/RS485)
Use the following parameters to configure the instrument for communication via the 8-pin M12 connection:
If the 9-pin D-sub (side) connector of the instrument is configured for RS485 communication (FLOW-BUS/Modbus), it will
not respond when connected to an RS232 configuration. In that case, use the power-up functionality of the
to switch to configuration mode and enable RS232 communication.
After configuring the required parameters, remember to return the instrument to the original communication mode.
Bus 2 Address
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…255
309
124/10
0x0F8A/3979
Bus 2 Baud Rate
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned long
RW
0…1.0E10
310
124/9
0xFC48…0xFC49/64585…64586
Bus 2 Parity
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…2
336
124/12
0x0F8C/3981
The following values are supported:
Value
Description
0
No parity
1
Odd parity
2
Even parity