MASS-STREAM
TM
Manual Flow Switch D-6300/FS
9.17.114A – edition: 2017-07
Page 36 of 36
Table 3: LED indications using micro switch
at normal running mode of an instrument
green
LED
red
LED
Time
Indication
off
off
0 – 1 sec
Pressing a switch shortly by accident will not cause
unwanted reactions of instrument
Pressing the switch 3 times briefly with intervals of
max. 1 sec. will force instrument to indicate its bus-
address/MAC-ID and maybe baud rate
off
on
4 – 8 sec
Reset instrument
Instrument program will be restarted and all
warning and error message will be cleared
During (new) start-up instrument will perform a
(new) self-test
on
off
8 – 12 sec
Auto-zero
Instrument will be re-adjusted for measurement of
zero-flow (not for pressure meter/controller)
NOTE: First ensure there is no flow and instrument
is connected to power for at least 30 minutes!
on
on
12 – 16 sec
Prepare instrument for FLASH mode
At next power-up FLASH mode will be active
This mode will be indicated by both LED’s off when
instrument is normally powered
Please note: “on“ = short flash (0.1 sec on / 2 sec off)
Table 4: LED indications using micro switch
at power-up situation of an instrument
green LED red LED
Time
Indication
off
off
0 – 4 sec
No action
Pressing a switch shortly by accident will not
cause unwanted reactions of the instrument
off
normal
wink
4 – 8 sec
Restore parameters
All parameter settings will be restored to
situation of final test at production
normal
flash
off
8 – 12 sec
Special mode for fieldbus applications –
Not relevant for Flow Switch instruments
normal
wink
normal
wink
12 – 16 sec
Special mode for fieldbus applications –
Not relevant for Flow Switch instruments
Please note: normal flash = 0.2 sec on / 0.2 sec off