3. PRODUCT OVERVIEW

MED0220GD-NR 
3736217-NR

MED0374GD-NR 
3736218-NR

Feature

Unit, Brand, Description

Data

Data

External Dimension

Height, mm

1310

1840

Width, mm

595

595

Depth, mm

598

598

Internal Dimension

Height, mm

1025

1555

Width, mm

505

505

Depth, mm

462

462

Capacity

Litres

215

372

Weight

Kg

63

81

Air Circulation

Fan-Forced

Fan-Forced

Climate Class

4

4

Temperature

Range

+2°C to 8°C

+2°C to 8°C

Min/Max Temp Recorder

Yes

Yes

Display

Digital

Digital

Exterior

 

Material

Powder coated sheet metal

Powder coated sheet metal

Electrical

Voltage, V

240

240

Frequency, Hz

50

50

Input Power, W

235

350

Current, A

1.18

1.23

Door

Solid or Glass

Low-E Glass + Argon Gas

Low-E Glass + Argon Gas

Shape

Flat

Flat

Self-Closing

Yes

Yes

Door Lock

Yes

Yes

Door Switch

Yes

Yes

Shelves

Qty

4

6

Type

Open rung shelf

Open rung shelf

Depth, mm

400

400

Adjustable

Yes

Yes

Capacity, kg

20

20

Internal Lighting

Internal Light

Yes

Yes

Alarms

Temperature

Yes

Yes

Audible

Yes

Yes

Compressor

Refrigerant

R600a

R600a

Charge, g

70

70

Defrost

Type

Off Cycle

Off Cycle

Manual/Automatic

Automatic

Automatic

Plug

Type

3-pin AU plug 10A

3-pin AU plug 10A

Feet/Wheels

Type

2 rollers at rear, 2 feet at front

2 rollers at rear, 2 feet at front

Summary of Contents for MEDIFRIDGE MEDO220GD-NR

Page 1: ...Confidence and control 08 06 2022 INSTRUCTION MANUAL MEDIFRIDGE DISPLAY FRIDGE MEDO220GD NR MEDO374GD NR...

Page 2: ......

Page 3: ...This Installation operation and service manual may not be reproduced in any form with prior written consent from Bromic Refrigeration 1 PREFIX 4 2 IMPORTANT NOTES WARNINGS 5 3 PRODUCT OVERVIEW 7 4 INS...

Page 4: ...ict con trol Refer to the National Vaccine Storage Guidelines Strive for 5 for more information on vaccine storage and handling or contact your local state of territory health department https www hea...

Page 5: ...ing the baskets All work on electrical equipment must be carried out by a qualified and licensed technician only Do not perform maintenance until the appliance has been turned off and power disconnect...

Page 6: ...frigeration circuit becomes damaged Avoid open flames and other source of ignition Immediately disconnect the appliance from power supply Ventilate the room and call Service Hotline WARNING Leaking re...

Page 7: ...ated sheet metal Powder coated sheet metal Electrical Voltage V 240 240 Frequency Hz 50 50 Input Power W 235 350 Current A 1 18 1 23 Door Solid or Glass Low E Glass Argon Gas Low E Glass Argon Gas Sha...

Page 8: ...ll the appliance where it is exposed to warm draughts or open doors Keep the air intake and outlets free of obstructions Adequate allowance should be made for door openings Ventilation Ensure the foll...

Page 9: ...e is rated 70 C It is recommended that the cable should not come into contact with parts over 50 C Do not turn off power before consulting the person responsible Mark the power source clearly so the r...

Page 10: ...ed in The electronic controller will display the current cabinet temperature after the unit has been connected to the power supply The LED indicates that refrigeration circuit is operating During the...

Page 11: ...he maximum and minimum temperatures two 2 times a day Recordings should be taken before the first door opening of the day MIN Temperature i Press the arrow key ii The Lo message will be displayed foll...

Page 12: ...nimum Limit door opening durations to less than 30seconds Wait 10 minutes for product and cabinet temperature to stabilise before opening door again Ensure products and materials are between 2 and 8 C...

Page 13: ...the products and transfer to a suitable location prior to cleaning the inside of the cabinet Do not use any aggressive or abrasive cleaning agents Never use a water hose or a high pressure jet to clea...

Page 14: ...ventilation C Wall clearances below minimum distance D Dirty condenser coil E Ice build up on evaporator coil A Relocate unit B Keep ventilation slots free C Adjust to minimum clearance D Clean E Per...

Page 15: ...e and promoting re use of the high quality materials employed used commercial refrigeration equipment must be collected for expert disposal and recycling Therefore observe the following disposal regul...

Page 16: ...orage guidelines strive for 5 BROMIC CANNOT BE RESPONSIBLE FOR THE LOSS OF OR DAMAGE TO STOCK All Bromic products are manufactured to the highest quality standard and are supported by our Extra Care S...

Page 17: ...xample 1 SERVICE REQUEST PRIORITY A B PRIORITY A SERVICE PRIORITY B SERVICE SERVICE CALL PLACED BEFORE MIDDAY SERVICE CALL PLACED AFTER MIDDAY When the unit is NOT performing its primary function such...

Page 18: ...NOTES...

Page 19: ......

Page 20: ...Y HEAD OFFICE SHOWROOM BROMIC REFRIGERATION 10 Phiney Place Ingleburn Sydney NSW 2565 Australia T 1300 276 642 For further product information bromicrefrigeration com au BROMIC REFRIGERATION 10 Phiney...

Reviews: