6.  PRODUCT LOADING AND STORAGE

Start-up

•  Clean the interior surface with a soft cloth and warm water before loading.
•  Position top and bottom shelves at marked locations to maintain correct air flow in unit.
•  Recommend placement of product on centre of shelves or evenly distributed on shelfs with 

minimum 30mm air gap between containers or use of perforated baskets.

•  Never stack containers and do not load unit excessively whereby air flow across the cabinet is 

restricted.

•  Do not load product above the top shelf load line marked on the side of the cabinet.
•  Do not block the evaporator fan.
•  Use cold mass (i.e. 300ml saline water bottles) on unused shelves and ensure containers are 

separated by 30mm on all sides. This is essential for maintaining correct storage temperature 

during door openings.

•  Avoid partial shelf loading, use cold mass on unused space
•  Use cold mass (i.e. water bottle) on unused shelves.
•  Keep refrigerator door openings to a minimum. Limit door opening durations to less than 

30seconds. Wait 10 minutes for product and cabinet temperature to stabilise before opening 

door again.

•  Ensure products and materials are between 2 and 8°C before loading – if there is a breach in the 

cold-chain, isolate the affected products. Do not use or discard until advice is given.

•  Some products are sensitive to ultraviolet (UV) light, please see storage guidelines from the 

manufacturer before removing from packaging.

•  Do not store food and other goods in the refrigerator.
•  Do not change controller settings. This is essential for maintaining correct storage temperature.
•  Maximum recommended shelf load is 20kg. 

NOTE: Bromic cannot be responsible for loss of stock or damage to stock. 

Please refer to the ‘National Vaccine Storage Guidelines: Strive for 5’ for more information on 

vaccine storage and handling or contact your local state or territory health department.

(https://www.health.gov.au/resources/publications/national-vaccine-storage-guidelines-strive-for-5) 

WARNING

: Failure to follow the guidelines below will result in cabinet 

temperatures below 2°C and above 8°C

IMPORTANT

: Run the refrigerator for at least 2 hours before loading.

Summary of Contents for MEDIFRIDGE MEDO220GD-NR

Page 1: ...Confidence and control 08 06 2022 INSTRUCTION MANUAL MEDIFRIDGE DISPLAY FRIDGE MEDO220GD NR MEDO374GD NR...

Page 2: ......

Page 3: ...This Installation operation and service manual may not be reproduced in any form with prior written consent from Bromic Refrigeration 1 PREFIX 4 2 IMPORTANT NOTES WARNINGS 5 3 PRODUCT OVERVIEW 7 4 INS...

Page 4: ...ict con trol Refer to the National Vaccine Storage Guidelines Strive for 5 for more information on vaccine storage and handling or contact your local state of territory health department https www hea...

Page 5: ...ing the baskets All work on electrical equipment must be carried out by a qualified and licensed technician only Do not perform maintenance until the appliance has been turned off and power disconnect...

Page 6: ...frigeration circuit becomes damaged Avoid open flames and other source of ignition Immediately disconnect the appliance from power supply Ventilate the room and call Service Hotline WARNING Leaking re...

Page 7: ...ated sheet metal Powder coated sheet metal Electrical Voltage V 240 240 Frequency Hz 50 50 Input Power W 235 350 Current A 1 18 1 23 Door Solid or Glass Low E Glass Argon Gas Low E Glass Argon Gas Sha...

Page 8: ...ll the appliance where it is exposed to warm draughts or open doors Keep the air intake and outlets free of obstructions Adequate allowance should be made for door openings Ventilation Ensure the foll...

Page 9: ...e is rated 70 C It is recommended that the cable should not come into contact with parts over 50 C Do not turn off power before consulting the person responsible Mark the power source clearly so the r...

Page 10: ...ed in The electronic controller will display the current cabinet temperature after the unit has been connected to the power supply The LED indicates that refrigeration circuit is operating During the...

Page 11: ...he maximum and minimum temperatures two 2 times a day Recordings should be taken before the first door opening of the day MIN Temperature i Press the arrow key ii The Lo message will be displayed foll...

Page 12: ...nimum Limit door opening durations to less than 30seconds Wait 10 minutes for product and cabinet temperature to stabilise before opening door again Ensure products and materials are between 2 and 8 C...

Page 13: ...the products and transfer to a suitable location prior to cleaning the inside of the cabinet Do not use any aggressive or abrasive cleaning agents Never use a water hose or a high pressure jet to clea...

Page 14: ...ventilation C Wall clearances below minimum distance D Dirty condenser coil E Ice build up on evaporator coil A Relocate unit B Keep ventilation slots free C Adjust to minimum clearance D Clean E Per...

Page 15: ...e and promoting re use of the high quality materials employed used commercial refrigeration equipment must be collected for expert disposal and recycling Therefore observe the following disposal regul...

Page 16: ...orage guidelines strive for 5 BROMIC CANNOT BE RESPONSIBLE FOR THE LOSS OF OR DAMAGE TO STOCK All Bromic products are manufactured to the highest quality standard and are supported by our Extra Care S...

Page 17: ...xample 1 SERVICE REQUEST PRIORITY A B PRIORITY A SERVICE PRIORITY B SERVICE SERVICE CALL PLACED BEFORE MIDDAY SERVICE CALL PLACED AFTER MIDDAY When the unit is NOT performing its primary function such...

Page 18: ...NOTES...

Page 19: ......

Page 20: ...Y HEAD OFFICE SHOWROOM BROMIC REFRIGERATION 10 Phiney Place Ingleburn Sydney NSW 2565 Australia T 1300 276 642 For further product information bromicrefrigeration com au BROMIC REFRIGERATION 10 Phiney...

Reviews: