Inspection 

 

Safety measures as described in Inspection/Maintenance schedule must be observed when 
inspecting or testing a chair. Only authorized caregivers or maintenance staff should attempt to 
inspect or test a chair. 
 
BRODA recommends regular visual inspection for overall signs of wear, damage, loose or 
missing fittings, and other safety concerns. If a breakage, defect, or operational problem is 
detected, the chair must be repaired inspected and tested for function before it is returned to 
service. 
BRODA recommends that the chair be inspected as often as each use if there is any reason to be 
concerned about the possibility of increased wear or loose or missing fittings.  
The visual inspection should be performed by the facility, or if in a private residence, by the 
individual responsible for the chair. 

 
5.2 

Functional Testing 

 

BRODA recommends that the Chair should be tested for operation of the chair’s functions 
without a resident in the chair.

 The testing may be as often as each use if there is any reason 

to be concerned about the possibility of increased wear or damage to the chair’s components.  
 
If the caregiver or maintenance department performing the functional testing believes that any 
function is not operating correctly, 

the chair should be taken out of service until a satisfactory 

functional test can be completed. 
 
The caregiver or maintenance department performing the testing should be aware that the 
seat tilt and leg rest elevation operations will be more difficult without a resident in the 
chair. 
 

When performing repairs or maintenance, do not use lubricants that contain solvents. Solvents 
will damage many of the moving components in the chair. If necessary, a white, food grade 
grease (lubricant) may be used on the sliding components in the chair, do not use spray 
lubricants on any part of the chair. 
 

WARNING:  
After any adjustments, repair or service, before use, make sure all attaching hardware is 
tightened securely.  Otherwise injury or damage may result. 
 
CAUTION: 
DO NOT over-tighten hardware as this could cause damage to the frame. 

 
Note: Please include the model number and serial number of the chair whenever you are ordering 
parts. These can be found on a sticker that is attached to the base on the bottom of the chair. 
 
Note:  Only a person with training on the BRODA Chair should attempt to do any repairs or 
maintenance on the chair.  Certain repairs may require a BRODA representative to personally 
attend to the repair.  If you are unsure as to what is required, call BRODA customer service at  
1-800-668-0637. 

Summary of Contents for LT tilt

Page 1: ...5 Operating Manual Distributor Rep This manual must be given to the caregiver responsible for this chair and its occupant Caregiver Before using this chair read this manual thoroughly and save for fut...

Page 2: ...WARNING 8 1 9 Accidental Motion Danger of Falling or Collision WARNING 8 1 10 Improper Restraint Use Risk of Serious Injury WARNING 8 1 11 Improper Use WARNING 8 1 12 Cleaning 9 1 13 Preventative Mai...

Page 3: ...est Angle Adjustable Independently Elevating Seat Back Surface BRODA Comfort Tension Seating Padding Removable Padding seat and back pad neckrest side panels legrest GENERAL INSTRUCTIONS The instructi...

Page 4: ...ions that are included in this manual are very important to the safe and effective operation of the BRODA Chair Safety requirements that are detailed in Section 1 must be followed at all times BRODA a...

Page 5: ...e professional Do not lean forward out of the chair any further than the length of the armrests Make sure the casters are pointing forward when leaning forward Otherwise the chair may tip over Do not...

Page 6: ...ly both to the first use of the Chair and to all subsequent uses An in service on the operation and safety issues as described in this manual must be given to the patient s caregivers prior to its fir...

Page 7: ...from pulling the chair over or pulling themselves off the seating surface and to prevent the patients from pulling moveable objects onto the chair and onto themselves We recommend that the chair be u...

Page 8: ...and exterior doorways Placing the patient near handrails tables or stationary objects while the caster locks are in place should be done with caution and should be supervised Failure to do so may lead...

Page 9: ...id and water Rinse with clean water and dry Step 2 Use a straight application of concentrated cleaners such as Formula 409 or Fantastik Spray Cleaner Then wipe with a clean cloth If using a hospital g...

Page 10: ...protected Leaving the chairs outdoors could cause the paint to peel Formula 409 is a trademark of the Clorox Company Fantastik is a trademark of DowBrands Inc Trademark of the Dow Chemical Company Ivo...

Page 11: ...nance do not use any lubricants that contain solvents Solvents will damage many of the moving components on the chair If necessary a white food grade grease lubricant may be used on the sliding compon...

Page 12: ...caregiver with the particular safety limits of the chair It is important that if assistance is required with the use of the chair that the caregiver remember to bend their knees and keep their back st...

Page 13: ...seat Transfer s refers to the movement of a resident into or out of a chair with the assistance of their caregiver s Mechanical Transfer s refers to the movement of a patient into or out of a chair w...

Page 14: ...upright condition and not stacked higher than 2 boxes tall No other materials should be shipped or stored on top of a BRODA box BRODA boxes should not be placed on pallets BRODA Chairs should be ship...

Page 15: ...local BRODA representative who supplies the chair or the HME Provider The caregiver s must adhere to the Safety Requirements at all times to ensure the safety of the patient caregivers and third parti...

Page 16: ...ting Seat Tilt 1 Squeeze the tilt handle which is located on the push bar The tilt will be easier to operate with a patient weight in the chair 2 Raise or lower the seat tilt until the desired angle i...

Page 17: ...r comes to an abrupt stop Check to ensure casters are clean and free of dirt Check to ensure all caster locks are functioning Check for loosening of the attachment between caster stem and housing Chec...

Page 18: ...functional testing believes that any function is not operating correctly the chair should be taken out of service until a satisfactory functional test can be completed The caregiver or maintenance de...

Page 19: ...uator housing 1 Loosen the nut on the bottom of the cylinder Unclip the cable from the actuator lever on the cylinder 2 Cut the cable ties that secure the cable to the frame of the chair carefully not...

Page 20: ...cated inside the barrel of the cylinder allowing for changes in position When the operator releases the handle the valve closes and locks the cylinder in the desired position The function of the actua...

Page 21: ...rom the vhair to make this adjustment 1 Loosen the lock nut located on the stem of the gas cylinder next to the actuator with a 17mm wrench 2 From the above instructions determine whether you want to...

Page 22: ...rts or warranty repairs The owner may return products to BRODA for warranty replacement or repair by shipping items prepaid and insured to the factory Warranty completed at the factory includes both m...

Reviews: