1.3
Cutaway of Operation
Model SU800 SingleSense™ Single-Use pH Sensor
Figure 2. Lever arm back and forward
September 2021 | 10
OM-062103 R1
Lever Arm Back
pH Sensor Retracted Into
Buffer Storage Chamber
The glass pH sensor is out of the process
media and measuring the buffer storage
chamber pH.
Lever Arm Forward
pH Sensor Inserted Into BPC
The glass pH sensor is inserted into the
process media and measuring the pH of
the process media.
Buffer storage
chamber
Process
media
Glass pH
sensor
O-ring
seal
1.0 How It Works
Operation Manual