3.3.1. Setting the Checkweighing Limits 

 

 

i. 

When the scale is ON, press 

FN 

(▲) key; 

 

UF-1 will be displayed. 

 

ii. 

Press

 FN 

(▲) key;

 

 

UF-2 will be displayed. 

 

iii. 

Press 

ON/ZERO/TARE 

(

) key; 

 

iv. 

To enter or edit the 

LO

 limit value, scroll through the 

displayed  digits  by  using  the 

ON/ZERO/TARE 

(

)

 

key.  Press  the 

FN 

(▲) key to increase the value of 

the flashing digit.

 

 

v. 

Press 

ON/ZERO/TARE 

(

)

 

key

 

to confirm; 

 

vi. 

Repeat points 

iv.

 and 

v.

 for the 

HI

 limit; 

 

vii. 

Choose  the 

beeper  mode  [000;  001;  002]

 

navigating  the  decimal  position  by  pressing 

ON/ZERO/TARE 

(

)

 

key.  Press  the 

FN 

(▲)  key  to 

increase the value of the flashing digit. 

 

 

viii. 

Press 

ON/ZERO/TARE 

(

)

 

key to complete and 

leave the procedure;

 

 

ix. 

Press 

UNITS 

(

ESC

)

 

key

 

to exit the User Menu. 

 

 

To disable the checkweighing feature, set the Hi and Lo limits to 0 in the User Menu [UF-
2].
 
The LO and HI annunciators will be turned off. 

 

3.3.2. Checkweighing Beeper Modes 

There are three modes. 

 

o 000 

The beeper is disabled.

 

o 001 

The beeper sounds only when the load is within the acceptable limits.

 

o 002 

The beeper sounds only when the load is outside the acceptable limits.

 

 

 

 

Please, ignore any other mode not listed above

 

3.4. Changing Units of Measure 

1.  With the scale turned on press the 

UNITS 

key to select the required unit of measure you want the 

scale to work in from kg 

– grams –lb. 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for C3236

Page 1: ...AWT35 501815 Issue AB C3236_u_en_501815 docx Page 1 User Manual Model C3236 Checkweigher...

Page 2: ...ntretien de routine 6 Cleaning the Machine Nettoyage de la machine 6 Sharp Objects Objets tranchants 6 2 Introduction page 7 Unpacking 7 General Installation Guidelines 7 Levelling the Scale 7 Battery...

Page 3: ...iptions UF 12 Supervisor Menu 13 Supervisor Menu Levels 13 Supervisor Functions Descriptions LF 14 Quick Calibration Menu 15 Quick Calibration Menu Levels 15 Quick Calibration Functions Descriptions E...

Page 4: ...AWT35 501815 Issue AB C3236_u_en_501815 docx Page 4 Manual revision history Current Issue Date Created Details of Changes Draft Jan 2019 New...

Page 5: ...tion manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to...

Page 6: ...operation and calibration Application and usage will determine the frequency of calibration required for safe operation PRUDENCE Le fonctionnement et l talonnage de cet quipement doivent tre v rifi s...

Page 7: ...ht or near air vents Place the scale on a level flat surface Do not place the scale near vibrating machinery Avoid unstable power sources Do not use near large users of electricity e g welding equipme...

Page 8: ...le please insure the power adaptor is fully removed from the scale prior to cleaning Figure 1 C3236 underside view 2 5 Front Panel Figure 2 shows the front panel of the unit It includes the display wi...

Page 9: ...n the scale The scale will self adjust to 0 Press this key to Zero the display Press this key to Tare the weight on the scale up to 100 of capacity In Menu mode It accepts the displayed item value in...

Page 10: ...the scale platform 4 Press the ON ZERO TARE key If the container weight is outside the Zero window area the display will show 0 000 weight and the NET annunciator is illuminated 5 Place the item to b...

Page 11: ...RE key Press the FN key to increase the value of the flashing digit viii Press ON ZERO TARE key to complete and leave the procedure ix Press UNITS ESC key to exit the User Menu To disable the checkwei...

Page 12: ...and increase a numeric value by pressing FN key Use ON ZERO TARE key to confirm and or move to the next option or numeric value position Press UNITS ESC key to exit the option or the menu 4 1 2 User...

Page 13: ...Values between 01 and 99 seconds are valid UF 5 ADC Update Rate use this to select the sampling frequency 15 1 30 2 or 7 5Hz 3 See Supervisor Menu UF 6 Automatic Zero Tracking use this to define a 0 5...

Page 14: ...a possible mechanical problem or possible load cell failure which linearity calibration may not be able to correct LF 4 ADC Update Rate use this to select the sampling frequency 15 1 30 2 or 7 5Hz 3 F...

Page 15: ...uick Calibration Menu Levels ECF Navigate the menus and increase a numeric value by pressing FN key Use ON ZERO TARE key to confirm and or move to the next option or numeric value position Press UNIT...

Page 16: ...to set the units range mode calibration units decimal position of the separator and division size Navigate the menus and increase a numeric value by pressing FN key Use ON ZERO TARE key to confirm and...

Page 17: ...few divisions If large inaccuracies are recorded this indicates a possible mechanical problem or possible loadcell failure which linearity calibration may not be able to correct LF 2 CALIBRATION SETT...

Page 18: ...number Description AWT25 502472 Power supply universal with USA UK EU plug adaptors 100 to 240VAC 50 60 Hz output 12V 1000mA with special water proof 2 pin connector Replacement battery DJW6 4 0 6V4...

Page 19: ...AWT35 501815 Issue AB C3236_u_en_501815 docx Page 19...

Reviews: