67XXU USB/Serial Scale Service Manual
81
9.2 Display Symbols
Symbol
Description
0----
Zero is over the setting range
0_____
Zero point is below the setting range
Ad----
Signal to ADC is over maximum range
Ad____
Signal to ADC is below minimum range
----
Weight is over upper limitation or display data is over limitation
____
Weight is below lower limitation
EEP.E1
CONFG or CAL parameters are not correctly set
EEP.E2
USER parameter is not correctly set
Lo.bAt
Battery voltage is lower than setting
CAP.- - -
Next displaying content is capacity
CAP.ER
Parameters about Capacity is not correct
CAL.Px
Calibration on point (x)
CAL.OFF
Calibration seal switch is on OFF position
CAL.ON
Calibration seal switch is on ON position
CAL.Er
Calibration error, maybe input data or loaded weight is incorrect, unstable, non-linear
CA.End
End calibration
OFF
Power OFF the indicator
STB.ER
Unstable time is larger than setting of USER-OTHER-NLD.RNG
PR.TAR
Preset TARE weight
COMP
Input COMPARE data mode
HIGH
Input HIGH limitation data of Comparison
LOW
Input LOW limitation data of Comparison
SPL.Lo
Sample load weight of low point.
SPL.HI
Sample load weight of high point.