52
67XXU USB/Serial Scale Service Manual
LINE
CAL.P0
Line calibration point0
Do zero point calibration. This point
can't be omitted.
CAL.P1
Line calibration point1
First weight point calibration. This
point can't be omitted and standard
weight must be over 10%FS.
END.Y
YES
End calibration?
YES=go to CAL.END to end
NO=go to do next point calibration
NO
CAL.P2
Line calibration point2
Second weight point calibration.
Standard weight must be over
10%FS and be larger than it in
CAL.P1. This point can be omitted.
END.Y
YES
End calibration?
YES=go to CAL.END to end
NO=go to do next point calibration
NO
CAL.P3
Line calibration point3:
Third weight point calibration.
Standard weight must be over
10%FS and be larger than it in
CAL.P2, this point can be omitted.
geo
code
00-70
Select geograhical position
Code 00-70
gravt
9.76183-
9.99999
Input gravity of user location
input
Input or view calibration
parameter values
Only used to copy calibration
parameters from one scale to a new
scale.
CAL
Submenu1
Submenu2
Option Parameter Description
Comment