31
Connecting auxiliary monitoring equipment
For monitoring or downloading data the Nippy jnr+ may be connected to a PC or
Laptop computer. The Nippy jnr+ isolated RS232 port is safe for use with any
domestic PC or laptop computer. However, when assembling a system, the
completed system should comply with EN60601-1 (medical systems). For
example, most computers do not comply with this standard, so it should be sited
at a distance, which makes it impossible to touch the computer and the patient at
the same time.
The RS232 port may also be used to run a remote alarm unit.