Guidance and manufacturer’s declaration – electromagnetic emissions
The ME equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of the ME equipment should assure that
it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The ME equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
Complies
Harmonic emissions
IEC 61000-3-2
Not Applicable
The ME equipment is
solely battery powered.
Voltage fluctuations/
flicker emissions
Not Applicable
Non-Life Support Equipment Separation Distance Calculation
(3Vrms / 3V/m compliance)
Rated maximum output
power of transmitter (W)
Separation distance according to
frequency of transmitter (m)
150 kHz to 80 MHz
in ISM bands
80 MHz to
800 MHz
800 MHz to
2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
d
= [
3,5
]
P
V
1
d
= [
3,5
]
P
E
1
d
= [
7
]
P
E
1