Unit Elements
4
Perfusor® compact, 2.1 gb
4 - 9
4.8 LS-Clip
Designation
Ord. No.
LS-clip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 7710
Exchange
1. Open unit
2. Press buzzer out of the holder.
3. Pull LS-clip out of the guide and exchange.
4. Assembly is performed in reverse order.
Setting the Loudness of the Alarm Tone (from serial number
38100 on)
1. Open battery compartment
.
2. Remove batteries.
3. Connect unit to mains power and switch unit on.
4. Disconnect from mains power for a short moment when the
switch-on test is finished. Pull out the mains power
connector and plug in again to trigger a device alarm (code
22, continuous tone).
5. Put a small flat blade screw driver (carefully) through the
battery compartment opening and set the volume desired.
6. Switch unit off via the keyboard.
7. Insert batteries.
8. Close battery compartment.
Fig.: 4 - 9
LS-clip
Fig.: 4 - 10
2.1
Summary of Contents for perfusor compact
Page 1: ...Perfusor compact Service Manual 0 10 3 1 0 Version 3 1 English...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes 2 1...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes 2 1...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes 2 1...
Page 56: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb 2 1 For your notes...
Page 62: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb 2 1 For your notes...
Page 64: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes 2 1...
Page 72: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb 2 1 For your notes...
Page 74: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes 2 1...
Page 78: ...Spare Parts List 10 10 4 Perfusor compact 2 1 gb For your notes 2 1...
Page 80: ...Revision Documentation 11 11 2 Perfusor compact 2 1 gb 2 1...
Page 85: ......