10
11
Changing the temperature scale
Your Braun ThermoScan is delivered to you with the Celsius (°C)
temperature scale activated. If you wish to switch to Fahrenheit (°F),
proceed as follows:
Make sure the thermometer is turned off.
Press and hold down the «start» button.
After about 8 seconds the display will
show «°C» flashing.
Release «start» button and press again
and release. The display will show «°F»
flashing.
After about 5 seconds the display will
auto reset and there will be 2 short beeps.
Replacing the battery
The thermometer is supplied with one 3V
Lithium cell battery (battery type). Insert new
battery when the low battery symbol appears
on the display.
Open the battery compartment. Remove the
battery by inserting the battery cover click into
the hole below the battery. Slide battery door
back until it snaps in place.
To protect the environment, dispose of empty
batteries at appropriate collection sites
according to national or local regulations.
Trouble-shooting
Situation
Solution
No lens filter is
attached.
Attach new, clean lens filter.
Measurement before
device stabilization
Wait until all the icons stop
flashing.
Ambient temperature
is not within the
allowed operating
range (50 – 104 °F or
10 – 40 °C).
Allow the thermometer to
remain for 30 minutes in a
room where the temperature
is between 50 and 104 °F
or 10 and 40 °C.
HI = too high
LO = too low
Temperature taken is not within
typical human temperature range
(93.2-108°F or 34-42.2°C)
Make sure the probe tip and
lens are clean and a new,
clean lens filter is attached.
Make sure the thermometer
is properly inserted. Then,
take a new temperature.
System error
(Er5 - Er9)
Reset the thermometer
by removing the battery and
then replacing it again, wait 1
minute, then turn on again.
Please contact your local Kaz
authorized service center.
(See page 13)
Battery is low.
Insert new battery.
Battery is too low.
Device cannot be
powered on to the
ready state.
Insert new battery.
Do you have any further
questions?
Please contact your local Kaz
authorized service center (See
page13 and Guarantee Card).
Summary of Contents for IRT 3020
Page 1: ...1 IRT 3020 ThermoScan Compact ear thermometer IRT 3020...
Page 8: ...14 15 1 2 3 4 6 5 7 start 1 2 3 4 start 5 6 7 10 40 C 50 104 F 20 C 50 C 95 30...
Page 10: ...18 19 1 2 3 Therm 20 5 start 4 6 start 4 60...
Page 11: ...20 21 LF 40 3 5 10 40 C 50 104 F 10 40 C 50 104 F 30 HI LO 34 42 2 C 93 2 108 F 5 9...
Page 15: ...28 29 1 2 2 1 3 start 4 4 34 6 C 36 8 C 5 start 4 6 start 4 60...
Page 16: ...30 31 1 2 3 Therm 20 LF 40...