Non-life support equipment separation distance calculation
(3Vrms / 3V/m compliance)
Rated maximum output
power of transmitter (W)
Separation distance according to
frequency of transmitter (m)
150kHz to 80MHz
in ISM bands
d = [
3,5
] P
V
1
80MHz to
800MHz
d = [
3,5
] P
E
1
800MHz to
2 .5GHz
d = [
7
] P
E
1
0 .01
0 .12
0 .12
0 .23
0 .1
0 .37
0 .37
0 .74
1
1 .17
1 .17
2 .33
10
3 .69
3 .69
7 .38
100
11 .67
11 .67
23 .33
Guidance and manufacturer’s declaration – electromagnetic immunity
The BUA5000AR is intended for use in the electromagnetic environment specified below .
The customer or the user of the BUA5000AR should assure that it is used in such an environment .
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
Complies
Floors should be wood,
concrete, or ceramic tile . If
floors are covered with synthetic
material, the relative humidity
should be at least 30% .
Radiated RF
IEC 61000-4-3
3V/m 80MHz to
2 .5GHz
Complies
Field strengths outside the
shielded location from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than 3V/m .
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Separation distance
calculation provided above . If
a known transmitter is present
the specific distance can be
calculated using the equations .
Conducted RF
IEC 61000-4-6
3Vrms 150kHz to
80MHz
Not
applicable
(no electrical
cabling)
Electrical fast
transient
IEC 61000-4-4
±2kV
power line
±1kV I/O lines
Not
applicable
The BUA5000AR
is solely battery powered .
Surge
IEC 61000-4-5
±1kV
differential
±2kV common
Not
applicable
Power frequency
magnetic field
IEC 61000-4-8
3A/m
Complies
Power frequency
magnetic fields should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment .
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
>95% dip 0 .5 cycle
60% dip 5 cycles
70% dip 25 cycles
95% dip 5 sec .
Not
applicable
The BUA5000AR
is solely battery powered .