Flow 50 Insufflator

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

On start-up - First screen offers surgical specialties 

 

Select appropriate 

Indication

” 

- f

ollowed by ”Veress” or “High Flow” 

to move to main screen.  

(Cardiac mode has 2 functions - 

“Vessel Harvesting” or “Field Flooding”

 (Field flooding is flow only)

 

 

 

 

 

 

 

 

Pre-set pressure can be set up to 30 mmHg 

Nb. 15 mmHg is set as the Safety Threshold “

Safety Threshold

” will flash repeatedly

 

Press 

+

 when the safety threshold message has gone off, to take pressure beyond 15 mmHg 

 

Press and hold the Flow 

+

  or 

 

buttons to go between “Low Flow”, Medium Flow” and “High Flow”

 

Low Flow = 3 L/min    Medium Flow = 20 L/min    High Flow = 40 L/min 

 

Insufflation is started and stopped using the “Start” button (Bottom Right)

 

(When insufflati

on is active, button displays “Stop”.  Insufflation is stopped by pressing “Stop”

 

 

CO

2

 volume delivered is reset by pressing the reset button 

Optional Smoke 

Evacuation Filter 

Housing 

Active Smoke Evacuation 

Actual 

Actual 

Volume 
& Reset 

System Settings 

Insufflation 

Gas Outlet 

Heated Gas 

Element 

(Optional) 

Start / Stop 
Insufflation 

Smoke Evacuation 

Footswitch 

(Optional) 

Pressure & Flow  

Pre-sets 

On / Off 