AESCULAP
®
3D
Einstein
Vision
® 3.0
Quick System Guide
32” 3D
Monitor
21” 2D
Keyboard
50L Insufflator
LED Light Source
3D / 2D CCU
Digital Documentation
3D Scope / Camera Holder
2D Camera Holder
CO
2
Gas Cylinder
Storage
A4 Thermal Sublimation
Printer
Page 1: ...on 3 0 Quick System Guide 32 3D Monitor 21 2D Monitor Keyboard 50L Insufflator LED Light Source 3D 2D CCU Digital Documentation 3D Scope Camera Holder 2D Camera Holder CO2 Gas Cylinder Storage A4 Ther...
Page 2: ...ht side of 21 monitor At the bottom of the monitor 4 Plug in integrated 3D light lead 5 if using 3D scope Plug in 3D camera head 6 if using 3D scope Set Insufflator to desired pressure 7 and flow rate...
Page 3: ...ne off to take pressure beyond 15 mmHg Press and hold the Flow or buttons to go between Low Flow Medium Flow and High Flow Low Flow 3 L min Medium Flow 20 L min High Flow 40 L min Insufflation is star...
Page 4: ...irmly into light aperture Do not rotate cable release Rotate clockwise to increase light output Press light source control to activate Deactivate light output Rotate anti clockwise to decrease light o...
Page 5: ...ds Camera White Balance can be performed from camera head remote button Rt or from CCU WB button Camera head button Long press white button 2sec CCU WB Button Short press 2sec Long press of CCU white...
Page 6: ...AESCULAP 3D EinsteinVision 3 0 2D 3DCamera Heads 3D Integrated Camera Head Button Configuration 2D Camera Head Button Configuration...
Page 7: ...MAKE ANY CHANGES TO THE MONITOR SETTINGS There is a 3D 2D button on the monitor DO NOT USE THIS BUTTON to swap between 3D and 2D Using the 3D 2D button on the monitor to select 2D effectively renders...