Page 31
Manual Operating Instructions
Release Valve
maximum
30 inch lbs
minimum
15 inch lbs
seats
(stops)
O
P
E
N
C
L
O
S
E
approximate
1/16" intervals
Valve Tightening
Specification:
Once valve seats
(stops), tighten 15
to 30 inch pounds
as shown.
Release Valve
Note:
Close backup pump release valve
securely
before
operating electric pump.
A
lift operation procedures (such as
loading and unloading passen-
gers).
Follow all Lift Operation
Safety Precautions!
Remove the pump cover to gain
access to the pump handle and
the hand pump. See photos on
page 30. To remove the pump
cover, turn the wing nut located
on top 1/4 turn and lift the pump
cover off.
The pump handle is
inside this cover secured by two
clips.
Rotate the top clip to
remove the pump handle.
Hand Pump