6a. Specifications 

Model

Gear Ratio

Max. Rotation Speed (Motor)

Max. Rotation Speed (Bur)

Spray

Glass Rod Optic

Bur Type

Coolant Delivery

Forza F5X

1:5 Increasing

40,000 RPM

200,000 RPM

Four Port Spray

Non-Optic

ISO 1797-1, Type 3 (Ø1.59 - 1.60mm Standard Bur)

External

Summary of Contents for Forza F1

Page 1: ...Contra Angle Attachments OPERATION MANUAL Please read this Operation Manual carefully before use and file for future reference...

Page 2: ...or disconnect the bur and or handpiece until the motor has completely stopped Before use always confirm that the bur is correctly seated and locked into place Do not use bent damaged or sub standard b...

Page 3: ...not use burs longer than 26 mm Its shank may warp under load at a high speed and spin out or lead to premature bearing failure Entering foreign object into gear or inside of handpiece head may cause b...

Page 4: ...Mounting Align the handpiece to the motor and push together For F5 and F1 optic Locate the handpiece and alignment pin straight into positioning of the motor 2 Removing Hold the motor and the handpie...

Page 5: ...econds until the oil comes out of the handpiece head 2 Lubrication Fig 1 E type Spray Nozzle Handpiece Spray Spray Cautions on Lubrication Be sure to hold the handpiece firmly to prevent the handpiece...

Page 6: ...not use a wire brush 2 Lubricate with spray 3 Insert into an autoclave pouch Seal the pouch 4 Autoclavable up to max 135 C ex Autoclave for 20 min at 121 C or 15 min at 132 C 5 Keep the handpiece in t...

Page 7: ...the set value Do not autoclave the handpiece even when it is bagged together with other instruments from which chemicals are not sufficiently cleaned This is to avoid discoloration and or damage to t...

Page 8: ...ular Glass Optic Rod entry point and exit point with an alcohol immersed cotton swab Remove all debris and oil Fig 2 Caution Do not use a sharp tool to clean the Cellular Glass Optic Rod It could dama...

Page 9: ...5...

Page 10: ...ns Model Gear Ratio Max Rotation Speed Motor Max Rotation Speed Bur Spray Glass Rod Optic Bur Type Forza F5 1 5 Increasing 40 000 RPM 200 000 RPM Four Port Spray Glass Rod ISO 1797 1 Type 3 1 59 1 60m...

Page 11: ...g the Bur 1 Depressing the push button 2 Insert the bur into the chuck Fig 3 3 Make sure it inserted firmly till the end of the head Fig 3 Removing the Bur 1 Press the button 2 Pull the bur from the h...

Page 12: ...6...

Page 13: ...Ratio Max Rotation Speed Motor Max Rotation Speed Bur Spray Glass Rod Optic Bur Type Coolant Delivery Forza F5X 1 5 Increasing 40 000 RPM 200 000 RPM Four Port Spray Non Optic ISO 1797 1 Type 3 1 59...

Page 14: ...ing Bur Mounting the Bur 1 Insert the bur into the chuck Fig 5 3 Make sure it inserted firmly till the end of the head Fig 5 Removing the Bur 1 Press the button 2 Pull the bur from the head Fig 6 Fig...

Page 15: ...6c Irrigation Nozzle Irrigation Nozzle Irrigation Tube Handpiece Fig 7 When irrigation is required please connect the irrigation tube to the irrigation nozzle F 5 X Motor...