Supply spray after each use and/or before autoclaving.
1) Mount the E-type spray nozzle over the spray nozzle until it firmly seats.
2) Shake the can 3-4 times to well mix lubricant and propellant.
3) Insert the E-type spray nozzle in the rear of the handpiece and spray for
approximately 2-3 seconds until the oil comes out of the handpiece head.
2. Lubrication
Fig.1
E-type Spray Nozzle
Handpiece
Spray
Spray
Cautions on Lubrication
• Be sure to hold the handpiece firmly to prevent the handpiece from slipping out by the spray pressure when lubricating.
• Supply lubricant until it comes out of the handpiece head (for approx. 2 seconds).
• Keep the Brasseler’s spray can upright.
• Use of Lubricants other than Brasseler’s Spray can cause excessive heat generation and possible handpiece damage.