13
7.4
Flow Gas Flow Meter Technical Characteristics
G
ENERAL
Measuring resolution: 10 ml
Measuring range: 20 to 4000 ml/h
Signal output for flow rate: 4-20 mA
Repeatability: 1%
Working principle: liquid displacement & buoyancy
Built-in sensors: pressure and temperature
LCD display: normalised gas flow (NmL/h), volume (Nml) and time (min)
Materials: high quality aluminium
Dimension: 19×14×6 cm
Weight: 0.180 kg
Power supply: input 100-240 V
50/60 Hz, output 12 V DC / 0.5 A
Usage: indoor applications for inert and slightly aggressive gases
F
EATURES
Large linear detection range (up to 4000 ml/h), which fits perfectly with most of lab-
and small pilot-scale applications
Embedded timer and reset button for easy logging of accumulated gas volume over a
defined period of time
Automatic gas flow and volume normalisation with real-time temperature and pressure
compensation, which allow for gas measurement standardisation
Standard analogue current signal output (4-20 mA) for automatic data acquisition
Adjustable foot and horizontal gauge to ensure the correct installation position for high
precision measurements
Glass window for visualisation of flow cell movement
Low maintenance requirements
Pre-calibration to ensure measurement accuracy
Summary of Contents for uFlow
Page 1: ...µFlow Gas flow meter Operation and Maintenance Manual ...
Page 2: ......
Page 3: ...µFlow ...
Page 26: ...21 2 Open the Volume folder 3 Select the setup exe file and open it ...
Page 28: ...23 5 Click next ...
Page 29: ...24 6 When the installation summary shows up start the installation by clicking Next ...
Page 30: ...25 7 A window will show the installation progress ...
Page 32: ...27 11 Open the Volume folder 12 Select the setup exe file and open it ...
Page 33: ...28 13 The installation is started and a green status bar shows the progress ...
Page 34: ...29 14 Select a destination directory for the files ...
Page 35: ...30 15 Start the installation by clicking Next ...
Page 36: ...31 16 The green status bars show the progress ...