7
SCHM-001
Rev 2
General Product Information
Frame
Welded steel with powder coated finish
Upholstery
Standard upholstery is vinyl
Upgrades are available
Movement
LC Base:
Lift Column - 15.81” (402mm)
Back:
Electric Actuator
Tilt:
Electric Actuator
Toe:
Electric Actuator
Standard
Swivel
Mobile
Net Weight
S2601:
334 lbs (151.5 kg)
372 lbs (168.7 kg)
340 lbs (154.2 kg)
S2605:
339 lbs (153.8 kg)
377 lbs (171.0 kg)
345 lbs (156.5 kg)
Electrical
24V D/C low voltage motors
Microprocessor controlled
Chair operates at 24V D/C
Power Supply
115VAC 60Hz 10A
230VAC 50-60 Hz 7A
(Refer to Serial Number Sticker for required voltage)
Protection
Class: I
Electrical Shock
Type: B
Liquid Protection
Ordinary equipment (IPXO)
Flammable Gases
Not suitable for use in the presence of flammable gases or where such gases may accumulate in
concentration. (Closed Space)
Operation Mode
Intermittent duty cycle:
10% - 2 minutes On / 18 minutes Off
Indoor use only
Operating
Temperature: 5
°
C to 40
°
C (41
°
F to 104
°
F)
Temperature
Relative Humidity: 20% to 90% @ 30
°
C - Not condensing
Atmospheric Pressure: 97kPa
Storage / Transportation
Temperature: -10
°
C TO 50
°
C (14
°
F TO 122
°
F)
Temperature
Relative Humidity: 20% to 90% @ 30
°
C - Not condensing
Atmospheric Pressure: 45 to 115kPa
Weight Capacity
500 lbs (228 kg) Subject to normal weight distribution
Note: Do not sit on back or adjustable toe section. These sections are not designed to support the full
weight of patient. Do not lift or slide chair using back section or seat upholstery.
Warning:
Do not modify this equipment without authorization from Boyd Industries, Inc.