Boyd S2601 Product Information Manual Download Page 13

13 

 

 

SCHM-001 

Rev 2 

Model: S2601 & S2605 Chairs 

Programming Instructions: 

  
1.  Read these instructions completely before you begin programming. 
  
2.  Using the motor controls, position the chair as desired. 
  
3.  Wait at least three (3) seconds and push and hold the save (S) button for 

four (4) seconds.  The system will issue a confirmation beep.  Within two 
(2) additional seconds, push the button of the desired position (i.e. 
“1”,”2”,”3”). The system will issue a second confirmation beep.  This will 
automatically replace any pre-programmed position. 

  
4.  Reposition chair and recall that position to ensure that it has been 

programmed correctly. 

  
5.  Repeat steps 2, 3, and 4 until all desired positions are programmed. You 

can save up to three (3) positions. 

  
6.  HOME is a preset position that cannot be changed. 
  
The programmed positions can be attained at any time and from any position. 
  

General Operating Instructions: 

  
1.  Before operating chair, watch for any obstructions under chair. 

 

Do not allow small children to play on, around, or underneath 
chair or base. 

 

Do not allow service animals on or underneath chair or base. 

 

Any obstruction that may hinder normal movement of any part of 
the chair should be cleared away before initializing any 
positioning of the chair. 

 

Check to be certain that all people are clear of the chair before 
operating chair functions. 

  
2.  Chair should be turned OFF at main power switch (C1) any time it is not 

being used. 

Operating Instructions 

Outlet for use with Boyd Light Accessory 
Only and may not be repositioned.  
Improper use could lead to injury or 
equipment damage. 

Foot Control 

 

Save 

Position 2 

Position 3 

Position 1 

230V 
Outlet 

115V 
Outlet 

Outlet Location 

Summary of Contents for S2601

Page 1: ...2 Product Information Manual S2601 S2605 Surgery Chair 12900 44th Street North Clearwater Florida 33762 800 255 2693 727 561 9292 FAX 727 561 9393 SALES FAX 727 573 1682 SERVICE www boydindustries co...

Page 2: ...Product Information Product Specifications Transport Disposal Unpacking and Power Requirements Installation Instructions Chair Install Initialization Procedure Operation Instructions Programming Inst...

Page 3: ...roduct is used in any way other than the product s design intent An authorized service technician must perform all service and maintenance Any warranty labor must be first approved by Boyd Failure to...

Page 4: ...4 SCHM 001 Rev 2 Safety Warning Symbols Caution Informational Read Technical Manual Before Servicing Main Power On Main Power Off Protective Earth Ground Type B Equipment General Warning Symbol Light...

Page 5: ...quipment or push objects of any type into openings as they could touch moving parts or dangerous voltage points inside equipment which could cause fire hazard electric shock and or bodily injury 13 Se...

Page 6: ...Controls Hand Controls are optional and connected using a coiled cable Pressing any motor control function switch will cancel any automatic movement function 3 Foot Controls Foot Control is attached w...

Page 7: ...PXO Flammable Gases Not suitable for use in the presence of flammable gases or where such gases may accumulate in concentration Closed Space Operation Mode Intermittent duty cycle 10 2 minutes On 18 m...

Page 8: ...to top of seat frame 35 03 to top of seat frame 24 86 62 68 32 75 with optional rail mount arms 62 68 SIDE VIEW S2601 S2605 Standard Lowered TOP VIEW S2605 TOP VIEW S2601 Raised Note Headrest height i...

Page 9: ...to top of seat frame 20 31 to top of seat frame 20 13 to top of seat frame 35 94 to top of seat frame SIDE VIEW S2601 S2605 Mobile Chair Lowered Raised Lowered Raised SIDE VIEW S2601 S2605 Swivel Chai...

Page 10: ...per lifting techniques to avoid physical injury Chair should be secured to transporting equipment and or transporting vehicle Transporting and Storage Temperatures for chairs to be kept between Temper...

Page 11: ...ed with upholstery Headrest Four 4 anchors Used to secure to floor A C power cable Accessories Options I E Arms Light post etc Owner s Operator s manual Note Accessories may be packaged separately Pow...

Page 12: ...ain power switch ON C1 located on the electrical panel at the front of base 6 Perform initialization procedure see below 7 Perform safety and operational check 8 If chair does not operate correctly se...

Page 13: ...ME is a preset position that cannot be changed The programmed positions can be attained at any time and from any position General Operating Instructions 1 Before operating chair watch for any obstruct...

Page 14: ...OSITION 1 Press to move to position 1 See separate instructions POSITION 2 Press to move to position 2 See separate instructions POSITION 3 Press to move to position 3 See separate instructions HOME R...

Page 15: ...st support the headrest with one hand and turn the threaded release knob direction shown below with the other Lower or raise the headrest to desired height Tighten knob direction shown below to lock i...

Page 16: ...ill lock arm board in set position Arm Board Height Rail Adjustment and Removal 1 Loosen the Arm Board Height and Rail Adjustment Handle to unlock arm board position See Below 2 The arm board can be r...

Page 17: ...on contact It is recommended that the chair and equipment be thoroughly cleaned and disinfected in accordance with the appropriate FDA guidelines Use only cleaners and disinfectants approved for clean...

Page 18: ...he Seat Upholstery Seat Upholstery Place the two Angle Brackets near the rear of the Seat Upholstery into open area shown below of Seat Frame Lower the upholstery so that the Locking Angle near the fr...

Page 19: ...e Parts List that they are completely engaged 3 Press and hold both position 1 position 2 switches down simultaneously for five to ten seconds or until beeping stops and changes to a steady tone Depre...

Page 20: ...20 SCHM 001 Rev 2 Boyd Industries Inc 12900 44th Street N Clearwater FL 33762 U S A...

Reviews: