33
9.2 Cleaning
The device should be cleaned on a regular basis, and the frequency of cleaning should be
increased in the areas suffering from serious environmental pollution or heavy sand.
Please consult or learn about the hospital’s regulations on the device cleaning before
cleaning.
The following cleaning agents are available for selection:
Diluted soapy water and diluted ammonia
Sodium hypochlorite (bleaching power for washing)
Hydrogen peroxide (3%)
Ethanol (70%)
Isopropanol (70%)
When cleaning the device:
1.
Turn off the oximeter.
2.
Use a soft cotton ball and adsorb the right amount of cleaning agent, and wipe the
display screen.
3.
Use a soft cloth ball and adsorb the right amount of cleaning agent, and wipe the
surface of the device.
4.
Use a dry cloth to remove excessive cleaning agent when necessary.
5.
The device is placed in the ventilated, shade and dry environment.
9.3 Disinfection
Disinfection operation may produce some degree of damage to the oximeter. It is
recommended to perform disinfection only when deemed necessary in your hospital
maintenance plan. The device should be cleaned before disinfection.
The recommended disinfectants are: 70% ethanol, 70% isopropyl alcohol and 2%
glutaraldehyde solution.