Before cleaning the device, you must turn off the power supply and disconnect
the power line and charger base.
Caution
If you pour liquid onto the device or accessories due to incaution, please
contact the maintenance personnel or the company immediately.
Attention
To clean or disinfect reusable accessories, please refer to the manual provided
along with the accessories.
9.1 Inspection
Before the oximeter is used for the first time, and after it is repaired or upgraded, or at
least every two years, a comprehensive inspection on it should be conducted by qualified
maintenance personnel to ensure the normal operation and work of the oximeter.
Inspection items should include:
The environment and power supply meet the requirements.
The device and accessories are not mechanically damaged.
The power line is not worn, with good insulation properties.
Use the specified accessories.
The alarm system functions properly.
Battery performance.
The monitoring function is in good working condition.
If you find any damage or unusual phenomenon, please do not use the oximeter, and
immediately contact the hospital's medical engineer or the company’s maintenance
personnel.