8 Reprocessing
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
41
8.6.
Inspection
These products are subject to wear when used as intended, depending on
the intensity of use. This wear arises from the design and construction of
the instruments and is unavoidable.
Replace the product if it shows externally visible defects or does not
function as described in this manual. Please advise the manufacturer or the
manufacturer’s authorised representative in such cases.
After cleaning, visually inspect each of the instruments and test them
for proper operation.
Replace any damaged parts.
DANGER
Risk of patient burns with brittle or defective insulation!
Replace instrument components with damaged
insulation.
Handle
1.
Check the actuation lever for ease of motion.
2.
Check the jaw holder for damage and corrosion.
Jaw
1.
Check the electrode surfaces and the blade groove for cleanliness
and damage.
2.
Check the cable in the joint area of the jaw for damage.
HF cable
1.
Check the connector for damage and corrosion.
2.
Inspect the insulation for damage.