• This equipment should not be used adjacent to or stacked with other equipment and that if

adjacent or stacked use is necessary, the Smoke Shark II should be observed to verify normal

operation in the configuration in which it will be used.

• Refer routine servicing to qualified biomedical technical personnel.

• Changes or modifications not expressly approved by Bovie Medical Corporation could void

the user’s authority to operate the equipment.

The warranty on this product is void if any of these warnings are disregarded. 

1.4.2    CAUTIONS:

• Federal law (United States of America) restricts this device to be used by, or on the order of a

physician.

• Do not block either the tubing or the filter. If either becomes occluded or significantly restrict-

    ed, the motor/blower may overheat and cause the unit to fail.

• Using any other filter or accessory not supplied by Bovie Medical Corporation may cause

damage and/or cause the system to be inoperable and may void the warranty.

• Care must be exercised in the installation of hoses, adapters and suction canisters.  Failure to

follow the procedures outlined in this manual may result in overheating of the motor and may

void the unit warranty.

• This device is not intended for evacuation of fluid.  If fluid is expected to be aspirated to the

Bovie® SF35 Filter, fluid collection devices must be installed with the vacuum hose assembly.

Failure to install a fluid collection device could cause filter blockage and electrical damage.

• The Bovie® SF35 should be changed according to the life of the filter.  The Bovie® SF35 Filter,

used with the Smoke Shark II Smoke Evacuation System(s), should not be used for more the

time specified for each filter. Failure to change the filter may result in decreased efficiency and

contamination of the electric motor, vacuum pump, and sound absorbing media within the

unit.

• Do not block either the tubing or the filter during operation. An occlusion or significant restric

tion may cause the motor to overheat and the unit to stop working.

• The installation of this equipment must be performed such that the intake and exhaust vents

located on the bottom of the system are not obstructed. Failure to properly install the unit may

cause reduced performance, damage and/or cause the system to be inoperable and may void

the warranty.

Summary of Contents for Smoke Shark II

Page 1: ...Operator s Manual 2020 Bovie Medical Corporation 5115 Ulmerton Road Clearwater Florida 33760 US Phone 1 800 537 2790 Fax 1 800 323 1640 International Phone 1 727 384 2323 www boviemed com...

Page 2: ...nts the Smoke Shark II against any defects in material or workmanship Bovie Medical Corporation will repair or replace at Bovie Medical Corporation s option the same without charge provided that routi...

Page 3: ...CTIONS 11 2 1 System Controls 2 2 Bovie SF35 Filter Instructions 2 3 Set up and Operation 2 4 Specifications 3 0 MAINTENANCE 20 3 1 General Maintenance Information 3 2 Cleaning 3 3 Periodic Inspection...

Page 4: ...S Patent 5874052 design captures particulates and micro organisms from 1 to 2 microns at an efficiency of 99 999 The third stage filtration uses the highest grade virgin activated carbon especially de...

Page 5: ...lowing systems 100 120 VAC 50 60 Hz 220 240 VAC 50 60 Hz 2 Type of protection against electrical shock Class I 3 Degree of protection against electric shock Type CF Applied Part 4 Degree of protection...

Page 6: ...cause harmful interference in which case the user will be required to correct the interference at their expense 13 This equipment operates in the following radio frequency specifications 14 To isolate...

Page 7: ...fety the tubing or wand should not come into direct contact with tissue Otherwise patient injury may result The Bovie Smoke Shark II filters and single use accessories are completely disposable Please...

Page 8: ...ers and suction canisters Failure to follow the procedures outlined in this manual may result in overheating of the motor and may void the unit warranty This device is not intended for evacuation of f...

Page 9: ...0 to 75 An atmospheric pressure range of 700 hPA to 1 060 hPa Storage environmental ambient temperature 14 F to140 F 10 C to 60 C Storage environmental relative humidity 10 to 75 There are no user ser...

Page 10: ...ROTECTIVE EARTH GROUND EQUIPOTENTIALITY DENOTES THE DATE THE EQUIPMENT WAS MANUFACTURED MANUFACTURER NON IONIZING RADIATION MANDATORY REFER TO INSTRUCTION MANUAL GUIDE DANGER EXPLOSION RISK IF USED WI...

Page 11: ...OUGH HIGH MEDIUM AND LOW SPEEDS MOTOR ON LED INDICATOR STANDBY MODE LED INDICATOR SERVICE TROUBLESHOOTING REQUIRED LED INDICATOR MOTOR SPEED LED INDICATORS Upper Light HIGH SPEED Center Light MEDIUM S...

Page 12: ...ne switch to illuminate yellow STANDBY LED Turn the system main power off by unplugging the power cord from the appliance inlet on the unit or the receptacle in the wall SUCTION CONTROL Membrane Contr...

Page 13: ...REEN LED indicating 100 filter life The indicator will progress thru subsequent GREEN LEDs to an AMBER LED as time elapses and begin flashing RED to indicate that the filter has expended its useful li...

Page 14: ...nging turn the smoke evacuation system off and disconnect any accessory tubing attached to the filter 2 Depress the filter tab and pull the filter from the smoke evacuation system and dispose of in ac...

Page 15: ...e tubing in such a way as to minimize potential trip hazard by users or patients 5 Activate the power unit by a Pressing the suction ON OFF switch on the membrane panel b Depressing and releasing the...

Page 16: ...2mm Standard Hose I D LPM Maximum Flow Setting lfm Metric 1 4 PERFORMANCE 7 8 Standard Hose I D CFM Maximum Flow Setting Model Number Motor Static Suction kPa 6 5 mm orifice 100 120 VAC 220 240 VAC 10...

Page 17: ...ble The model Smoke Shark II is suitable for use in all establishments other than domestic establishments and those directly connected to the public low voltage power supply network that supplies buil...

Page 18: ...user of the Smoke Shark II requires continued operation during power mains interrup tions it is recommended that the Smoke Shark II be powered by an uninterruptible power supply or battery 0 UT for 0...

Page 19: ...Smoke Shark II Operator s Manual Page 18 Single phase at 0 0 at UT 250 300 cycles Single phase at 0 0 at UT 250 300 cycles 30 A m 30 A m...

Page 20: ...hark II Table 3 Portable and mobile RF communications equipment should be used no closer to any part of the Smoke Shark II including cables than 30 cm 12 in Field strengths from fixed RF transmitters...

Page 21: ...r inlet Obvious external or internal damage to the system 3 4 Troubleshooting the System see below Maintenance Section 3 0 PROBLEM POTENTIAL CAUSE CORRECTIVE ACTION 1 Smoke Evacuation System is ON but...

Page 22: ...it to properly return your unit If you do not have the original packing material ask Customer Service for advice on how to pack the unit for return shipment Step 5 Freight for all returned goods shoul...

Page 23: ...demands of third parties of any kind relating to the system and its use SPECIFICATIONS Specifications are subject to change without notice SHIPMENT OF ORDER Bovie Medical Corporation will try to ac c...

Page 24: ...2020 Bovie Medical Corporation 5115 Ulmerton Road Clearwater Florida 33760 USA US Phone 1 800 537 2790 Fax 1 800 323 1640 International Phone 1 727 384 2323 www boviemed com...

Reviews: