Vercise™ DBS Programming Manual
91053785-17 30 of 1705
Vercise™ DBS Programming Manual
en
The two rates are dependent on one another in that only certain combinations of rates are
permitted. The table below indicates which frequencies are available for the second Area
at commonly used frequencies for the first Area.
First Area
Rate (Hz)
Subset of Available Second Area Rates (Hz)*
99
89
†
, 99, 112
†
, 113
†
, 116
†
, 119, 128
†
, 132, 139
†
, 149, 154
‡
,
159
†
, 170
‡
, 174
†
, 179
†
, 198, 223
†
, 231
†
, 238
‡
132
99, 113
†
, 116
‡
, 119
‡
, 132, 149
‡
, 154
‡
, 159
†
, 170
‡
, 176, 185
‡
,
198, 231
‡
, 238
‡
179
89, 99
†
, 104
‡
, 112
†
, 119, 128
†
, 139
‡
, 143
†
, 149
†
, 179, 208
‡
,
223
†
, 238
†
238
89
†
, 95, 99
‡
, 119, 132
‡
, 136
‡
, 143
†
, 149
‡
, 159, 170
‡
, 179
†
, 190
‡
,
198
‡
, 238
*Subset includes only frequencies in typical DBS range (89 to 255 Hz) using typical pulse
widths (30 to 90 μs).
†Available when both pulse widths are less than or equal to 60 μs.
‡
Available when both pulse widths are less than or equal to 30 μs.